User's Manual
  Wireless Service Manual 
  Page 114 of 124  RA14-11A-022 
Test ITEM (8) 
Resolution 
Frequency 
D: Daily  □  M: Monthly  ■ 
Q: Quarterly  □  A: Annually  □ 
Objective: 
Confirm the resolution of the detector. 
Equipment: 
Workstation, VXSetup S/W 
Power supply, Power Meter 
Resolution Chart (Nuclear Associate, model 07-523-2 or Line pair CHART 0.05mmPb CN37076) 
X-ray GEN., X-ray Tube 
Inspection Report Form 
Procedure: 
Turn on the power of the system. 
Run VXSetup. 
Open the Diagnose window. 
Attach resolution chart (Line pair CHART 0.05 ㎜ Pb CN37076, model 07-523-2) on the center of the 
detector with diagonal direction. 
Set X-ray generator to 50kVp, 2mAs and SID to 1m. 
Expose X-rays. 
Confirm that the resolution is over 3.5lp/㎜. 
Performance and Corrective Action: 
Confirm that the resolution is over 3.5lp/㎜. 
If the resolution is under 3.5lp/㎜, then test again by adjusting mAs from 1mAs to 5mAs. 
If still resolution is under 3.5lp/㎜, service assistance is necessary. 
Record result on Inspection Report Form. 
Result: 
Frequency: D: □,    M: □,    Q: □,    A: □  Date: 20**-**-**  Operator: 
Limit of Acceptability: 3. 5lp/㎜ 
Remarks: 










