  Page 117 of 124  RA14-11A-022 
Test ITEM (11) 
DQE 
Frequency 
D: Daily  □  M: Monthly  □ 
Q: Quarterly  ■  A: Annually  □ 
Objective: 
Evaluate the distinct characteristics of detector through the amount of radiation coming into Flat Panel 
Detector, resolution and contrast of images/projections, and the unification of noises of projection. 
Equipment: 
Workstation, VXvue S/W 
Power supply, Power Meter 
X-ray GEN., X-ray Tube 
Inspection Report Form 
Procedure: 
IEC62220-1(RQA 5) in accordance with the conditions 
Performance and Corrective Action: 
In case of CSI type, properties/qualities should be obtained more than 45% at 1l/p  ㎜. 
In case of Gadox type, properties/qualities should be obtained more than 30% at 1l/p  ㎜. 
If properties/qualities are not obtained, reset the conditions and test again. 
If properties/qualities are still not obtained after the second test, service assistance is necessary. 
Record result on Inspection Report Form. 
Result: 
Frequency: D: □,    M: □,    Q: □,    A: □  Date: 20**-**-**  Operator: 
Limit of Acceptability: CSI:  1l/p  ㎜, more than 45% 
  Gadox:    1l/p  ㎜, more than 30% 
Remarks: 