User's Manual
  Wireless Service Manual 
  Page 12 of 124  RA14-11A-022 
Notes for Using the Equipment 
System Diagnostic 
The VXSetup software runs a system diagnostic. Run VXSetup software after installing the system and at 
least once a year. If an error occurs, report the detailed error information to Vieworks local dealer or 
distributor. 
The owner is responsible for ensuring that the system diagnostic is performed every year. 
Do not try to use the system if the system diagnostic is failed. 
Calibration 
To ensure optimal performance of the system, it is important to verify that the system is calibrated. 
The owner is responsible for ensuring that the system calibration is performed after the 
system installation is completed or the system is repaired. Do not try to use the system if 
system calibration is not performed. 
Distances measurements 
Distances measurements in millimeters are possible only after distance calibration has been performed using 
a reference object (refer to VXvue User Manual). 
The operator is responsible for performing distance calibration with a reference object and 
verifying the results of the distance calibration before taking any distance measurements 
on an image. 
Left/Right Marker 
The operator is responsible for the correct and clear marking on the left or right side of the image to eliminate 
possible errors. 
The software includes an option to mark the image with L (left) or R (right) indicator from acquisition phase 
through printing and archiving. If the operator chose, for any reason, not to use L/R markers, he must use an 
alternative way to eliminate any possible mistake. 










