User's Manual
  Wireless Service Manual 
  Page 18 of 124  RA14-11A-022 
Immunity Test 
IEC 60601 
Test Level 
Compliance 
Level 
Electromagnetic Environment – Guidance 
Conducted RF   
IEC 61000-4-6 
Radiated RF   
IEC 61000-4-3 
3 Vrms 150 
㎑  to 80  ㎒ 
3 V/m 80  ㎒ 
to 2.5  ㎓ 
3 Vrms 150 
㎑  to 80  ㎒ 
3 V/m 80  ㎒ 
to 2.5  ㎓ 
Portable and mobile RF communications equipment 
should be used no closer to any part of the EUT, 
including cables, than the recommended separation 
distance calculated from the equation applicable to the 
frequency of the transmitter. 
Recommended separation distance 
Where P is the maximum output power rating of the 
transmitter in watts (W) according to the transmitter 
manufacturer and d is the recommended separation 
distance in meters (m). 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey
a
, should 
be less than the compliance level in each frequency 
range.
b
Interference may occur in the vicinity of equipment 
marked with the following symbol: 
NOTE 1: At 80  ㎒  and 800  ㎒, the higher frequency range applies. 
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by 
absorption and reflection from structures, objects and people. 
a
 Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and 
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted 
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an 
electromagnetic site survey should be considered. If the measured field strength in the location in which 
the EUT is used exceeds the applicable RF compliance level above, the EUT should be observed to 
verify normal operation. If abnormal performance is observed, additional measures may be necessary, 
such as reorienting or relocating the EUT. 
b
 Over the frequency range 150  ㎑  to 80  ㎒, field strengths should be less than [V
1
] V/m. 