User's Manual
  Wireless Service Manual 
  Page 40 of 124  RA14-11A-022 
2.6.2  Timing of Signals (DR Trigger Mode) 
A
A’
B B’
EXP_OK
EXP_REQ
Image Acquisition Time
Figure 2.12 Timing of Exposure signal 
  Image Acquisition Time+ 
  Exposure request signal A (EXP_REQ) should be applied first, then exposure respond signal B 
(EXP_OK). 
  Image Acquisition Time can be set from 40  ㎳  to 4,000  ㎳  with 1  ㎳  increment when default value 
is 500  ㎳. 
2.6.3  EXT_INF Port Pin Assignment 
5 1
15
11
20 16
30
26
Figure 2.13 EXT_INF port pin assignments 