User's Manual
  Wireless Service Manual 
  Page 90 of 124  RA14-11A-022 
Debounce_DR 
Debounce_DR is used to remove unwanted trigger signal occurred due to external noise when DR Trigger is 
set as Exposure Mode. It is recommended to use the Debounce_DR setting value as default 3  ㎳. We 
recommend setting the minimum exposure time of X-ray generator to more than 3  ㎳. If the value is set to 
less than 3  ㎳, the detector may not acquire images. 
Debounce_AED 
Debounce_AED is used to prevent unwanted imaging occurred due to external noise when AED is set as 
Exposure Mode. It is recommended to use the default values set by manufacturer. We recommend setting 
the minimum exposure time of X-ray generator to more than the value set in the Debounce_AED. If the value 
is set to less than the settings in the Debounce_AED, the detector may not acquire images. 
If unwanted images are acquired without X-ray exposure while operating the equipment, set Debounce_AED 
with 1  ㎳  increments. However, make sure not to exceed 5  ㎳  since an image degradation may appear in 
the image if Debounce AED exceeds 5  ㎳. 
5.4.1.3  Gain Type Setting 
To acquire X-ray image with proper brightness, adjust the Gain Type setting. You can select the Gain Type to 
adjust the sensitivity of the detector, then you can acquire X-ray images with desired brightness according to 
the specifications of X-ray generator or the type of objects. The default gain type is 1. 
The following table describes each Gain Type of the sensitivity ratio. 
Gain Type  0  1  2  3 
Gadox Detector  Sensitivity ratio  0.86 
1 
1.2  1.5 
CsI Detector  Sensitivity ratio  0.62 
1 
1.14  1.33 
The owner is responsible for ensuring that the Gain Pixel Correction is performed after 
adjusting the Gain Type. 