5.4.2  Offset Calibration 
5.4.2.1  Pre-offset Calibration 
1  Select Dark acquisition–Normal and click the Load pre-offset button to apply pre-offset data which is 
provided with the flat panel detector. 
2  When you create new pre-offset data, input the number of images to be acquired to Stage in the Pre-
offset calibration item and then click the Run pre-offset calibration button. 
3  Acquiring image process will be proceeded automatically, then the average value of acquired images 
will be displayed in Current Value and the current status will be displayed in Stage. 
4  After acquiring images, a window for saving created pre-offset data appears, then specify a directory to 
save the data. 