User's Manual
  Wireless Service Manual 
  Page 92 of 124  RA14-11A-022 
5.4.2.2  Post-offset Calibration 
1  Click the Get Post offset button to locally save the offset file that is backed up on the detector. 
2  When you create new post-offset data, input the number of images to be acquired to Stage in the Post-
offset calibration and click the Run post-offset calibration button. 
3  Acquiring image process will be proceeded automatically, then the average value of acquired images 
will be displayed in Current Value and the current status will be displayed in Stage. 
4  After acquiring images, a window for saving created post-offset data appears, then specify a directory to 
save the data. 
5  To save the previously created post-offset file to the detector, click the Upload post offset button and 
then select the file to upload. 
The offset data that is already uploaded to the detector will be used for acquiring images 
with wireless detector instead of the data stored in the local. 