VIVIX-S VW Service Manual © Vieworks. 2019 All rights reserved. The specifications and related information in this manual may be changed without notice. Refer to Vieworks Download System (VDS) for the latest version of our manuals. Under copyright laws, this manual should not be reproduced, in whole or in part, without the written permission of Vieworks.
VIVIX-S VW Service Manual Contents 1. Instruction ......................................................................................................................................... 6 1.1 Document Guide ..................................................................................................................................... 7 1.1.1 Caution .....................................................................................................................................................
VIVIX-S VW Service Manual 2.5 Battery Pack (FXRB-04A) ..................................................................................................................... 32 2.5.1 2.6 Specifications .................................................................................................................................................................32 SCU Lite (FXRP-02A) ...........................................................................................................................
VIVIX-S VW Service Manual 4.1 Product Installation .............................................................................................................................. 58 4.1.1 Connecting SCU mini (FXRS-04A) ........................................................................................................................58 4.1.2 Booting SCU mini (FXRS-04A) ...............................................................................................................................
VIVIX-S VW Service Manual 7.1.1 Product Lifetime ...........................................................................................................................................................87 7.1.2 Regular Inspection and Maintenance ................................................................................................................87 7.1.3 Repair ..................................................................................................................................
VIVIX-S VW Service Manual 1. Instruction This section gives basic information of this manual and products. Document Guide Intended Use Product Use Rev.1.
VIVIX-S VW Service Manual 1.1 Document Guide This operation manual explains how to use the VIVIX-S VW detector models made by Vieworks, X-ray interface unit, and other peripheral equipment. Use this manual to set up the VIVIX-S VW detector as well as understand its various functions. 1.1.1 Caution If the user is not fully aquainted with this manual, the product can be malfunctioned, or unsuspected problem can be happened due to carelessness.
VIVIX-S VW Service Manual 1.2 Intended Use VIVIX-S VW detectors are digital X-ray imaging solution. They acquire images by detecting X-rays that has been passed through the human body. When X-ray photons pass through the scintillator in the detector, the photons convert to visible ray, and the visible ray is converted to electronic signals through TFTs – thin film transistors (a-Si).
VIVIX-S VW Service Manual 1.3 Product Use This chapter provides instructions about the use of the product, disposal and the liability limit of Vieworks. 1.3.1 Product Usage 1 Only a physician or a legally certified operator should use this product. 2 The equipment should be kept in a safe and operable condition by maintenance personnel. 3 Follow the guidelines in this manual when installing and using this product. 4 Use only computers and image display monitors recommended in this manual.
VIVIX-S VW Service Manual 1.3.3 Product Disposal Disposal of this product in an unlawful manner may have a negative impact on health and on the environment. When disposing of this product, therefore, be absolutely sure to follow the procedure that conforms to the laws and regulations applicable in your area. European Union (and EEA*) only This symbol indicates that this product is not to be disposed with your household waste, according to the WEEE Directive (2012/19/EC) and your national law.
VIVIX-S VW Service Manual 2. Product This section gives an instruction about the product components and their specifications. Product Components FXRD-3643VAW, FXRD-3643VAW PLUS FXRD-4343VAW, FXRD-4343VAW PLUS FXRD-2530VAW, FXRD-2530VAW PLUS Battery Pack (FXRB-04A) SCU Lite (FXRP-02A) SCU mini (FXRS-04A) Others Rev.1.
VIVIX-S VW Service Manual 2.1 Product Components 2.1.1 Detector Product Name Image Model/Description ※ Basic: Battery Pack (2ea) FXRD-3643VAW (3.15kg) FXRD-3643VAW PLUS (3.15kg) ※ Option: Battery Pack 1ea FXRD-3643VAW (2.95kg) FXRD-3643VAW PLUS (2.95kg) ※ Basic: Battery Pack (2ea) FXRD-4343VAW (3.7kg) FXRD-4343VAW PLUS (3.7kg) VIVIX-S VW Wireless ※ Option: Battery Pack 1ea Detector FXRD-4343VAW (3.5kg) FXRD-4343VAW PLUS (3.5kg) ※ Basic: Battery Pack (1ea) FXRD-2530VAW (1.
VIVIX-S VW Service Manual La conception de la feuille de décoration fixée au détecteur peut être différente pour chaque client. La batterie est intégrée au détecteur. La batterie peut être remplacée en fin de vie. FXRD-3643VAW, FXRD-3643VAW PLUS, FXRD-4343VAW et FXRD-4343VAW PLUS sont équipés de deux batteries, dont une seule peut être installée en option sur demande. Les FXRD-2530VAW et FXRD-2530VAW PLUS ne peuvent être équipées que d’une seule batterie. 2.1.
VIVIX-S VW Service Manual 2.1.3 SCU (System Control Unit) Component Model / Description FXRP-02A (330g) DC Adapter SCU Lite (Option) BPM030S24F10 / DC 24V, Max 1.33A AC Power Cable (2m) 110V (1EA) / 220V (1EA) FXRS-04A (1.2kg) SCU mini (Option) AC-DC adapter BPM060S24F10 / DC 24V Max. 2.7A Rev.1.
VIVIX-S VW Service Manual AC Power Cable (2m) 110V (1EA) / 220V (1EA) SCU is not a standard item provided to customers. If necessary, the user can optionally purchase the desired SCU according to the inteded use. SCU Lite does not support wireless connections. Only use it as a wired connection. The SCU mini supports both wired and wireless connections. If you only need a wireless connection, you can use a separate external AP or detector AP.
VIVIX-S VW Service Manual 2.1.4 Accessories Type Image Description AC-DC adapter AC-DC Adaptor BM090S18F02 / DC 18V, Max. 5A Detector Operation manuals Detector Service manuals Hardware Manual CD Software files Software manuals Software / Software Manual CD (VXvue) Tether Interface Cable (7m) Generator Interface Cable (15m) Cables (Option) UTP LAN Cable (15m, Direct) CAT 5E or CAT6 (or higher) Rev.1.
VIVIX-S VW Service Manual Velcro tape (25mm x 50mm / 6ea) Attached to the corners of bucky and the friction surface of a detector. Velcro tape (25mm x 100mm / 6ea) The cable is not a standard item provided to customers. If necessary, the user can optionally purchase the desired cable according to the intended use. A tether interface cable is required for wired connections. Le câble n'est pas un élément standard fourni aux clients.
VIVIX-S VW Service Manual 2.2 FXRD-3643VAW, FXRD-3643VAW PLUS 2.2.1 Specifications Item Model Specifications FXRD-3643VAW FXRD-3643VAW PLUS Image Sensor TFT: a-Si (Amorphous Silicon) X-ray Scintillator Type FXRD-3643VAW : Csl type A FXRD-3643VAW PLUS : CsI type B Pixel Pitch 0.14㎜ (140㎛) Field of View 36cm x 43cm (14” x 17”) Active Area (H x V) 358.4㎜ × 430.08㎜ Active Array 2560 x 3072 pixels Effective Area 355.04㎜ x 426.
VIVIX-S VW Service Manual The longer you use the battery, the smaller its capacity. Plus vous utilisez la batterie, plus sa capacité est réduite. 2.2.2 Functions Side Name A B C D Tether Interface Connector Magnet for fixing the tether interface AC-DC Adapter Connector OLED Display Description Tether interface cable connector. Used for wired connection between a detector and SCU.
VIVIX-S VW Service Manual Rear Name Rev.1.0 Description The cover needs to be opened and closed when replacing A Battery Pack Cover B Handle A handle for carrying a detector C Lift Structure Used when the detector is placed on flat surface the battery pack.
VIVIX-S VW Service Manual 2.2.3 Deco Sheet Indication Info. Description A Center of the detector Indicates the central position of detector. B Image starting point Indicates the starting point of an original image. C~F Certification logo Indicates the certification logos relating to a medical device. Image starting point (0.0) of this detector is located nearby the tether interface connector.
VIVIX-S VW Service Manual 2.2.4 Wireless Communication Item Specifications Wireless standard IEEE 802.11n/ac 2.412 ~ 2.472㎓ (13 Channels) Frequency range 5.18 ~ 5.24㎓ (4 Channels) 5.745 ~ 5.805G㎐ (4 Channels) Data transmission rate Modulation Max. 300Mbps (IEEE802.11n, MIMO 2x2, 5GHz, 40MHz) Max. 1300Mbps (IEEE802.11ac, MIMO 3x3, 5GHz, 80MHz) BPSK, QPSK, 16-QAM, 64-QAM, 256-QAM Transmission power 2.4㎓: Max. 19dBm (per chain) 5㎓: Max.
VIVIX-S VW Service Manual 2.2.6 Load Limit of Detector Uniform load Local load Over the whole surface Center diameter 40 ㎜ Max. 300 kg Max. 150 kg Do not let the paitent or object heavier than load limit be on the detector. Then, detector can be damaged. Do not allow the patient to stand completely on the detector or lay down on it. Even weights within the load limits can seriously affect internal components such as sensors.
VIVIX-S VW Service Manual 2.3 FXRD-4343VAW, FXRD-4343VAW PLUS 2.3.1 Specifications Item Specifications Model FXRD-4343VAW FXRD-4343VAW PLUS Image Sensor TFT: a-Si (Amorphous Silicon) X-ray Scintillator Type FXRD-4343VAW: Csl A type FXRD-4343VAW PLUS: CsI B type Pixel Pitch 0.14㎜ (140㎛) Field of View 43cm x 43cm (17” x 17”) Active Area (H x V) 430.08㎜ × 430.08㎜ Active Array 3072 x 3072 pixels Effective Area 426.72㎜ x 426.
VIVIX-S VW Service Manual The longer you use the battery, the smaller its capacity. Plus vous utilisez la batterie, plus sa capacité est réduite. 2.3.2 Functions Side Name A B C D Tether Interface Connector Magnet for fixing the tether interface AC-DC Adapter Connector OLED Display Description Tether interface cable connector. Used for wired connection between a detector and SCU.
VIVIX-S VW Service Manual Rear Name Rev.1.0 Description The cover needs to be opened and closed when replacing A Battery Pack Cover B Handle A handle for carrying a detector C Lift Structure Used when the detector is placed on flat surface the battery pack.
VIVIX-S VW Service Manual 2.3.3 Deco Sheet Indication Info. Description A Center of the detector Indicates the central position of detector. B Image starting point Indicates the starting point of an original image. C~F Certification logo Indicates the certification logos relating to a medical device. Image starting point (0.0) of this detector is located nearby the tether interface connector.
VIVIX-S VW Service Manual 2.4 FXRD-2530VAW, FXRD-2530VAW PLUS 2.4.1 Specifications Item Specifications Model FXRD-2530VAW / FXRD-2530VAW PLUS Image Sensor TFT: a-Si (Amorphous Silicon) X-ray Scintillator Type FXRD-2530VAW: Csl A type FXRD-2530VAW PLUS: CsI B type Pixel Pitch 0.124㎜ (124㎛) Field of View 25cm x 32cm (10” x 12”) Active Area (H x V) 253.95㎜ × 317.44㎜ Active Array 2048 x 2560 pixels Effective Area 251.0㎜ x 314.
VIVIX-S VW Service Manual 2.4.2 Functions Side Name A B C Tether Interface Connector Magnet for fixing the tether interface AC-DC Adapter Connector Description Tether interface cable connector. Used for wired connection between a detector and SCU.
VIVIX-S VW Service Manual Rear Name Rev.1.0 A Battery Pack Cover C Lift Structure Description The cover needs to be opened and closed when replacing the battery pack.
VIVIX-S VW Service Manual 2.4.3 Deco Sheet Indication Info. Description A Center of the detector Indicates the central position of detector. B Image starting point Indicates the starting point of an original image. C~F Certification logo Indicates the certification logos relating to a medical device. Image starting point (0.0) of this detector is located nearby the tether interface connector.
VIVIX-S VW Service Manual 2.5 Battery Pack (FXRB-04A) 2.5.1 Specifications Item Specifications Model FXRB-04A Type Lithium Ion Polymer Nominal Voltage DC +11.55V Nominal Capacity 3,400mAh Number of Cells 3S1P (3 Series 1 Parallel) Approx. 800 times Battery Life (One time standard: Fully charged and then fully discharged) Dimensions (H × W × D) 189.0㎜ × 89.0㎜ × 6.65㎜ (max.) Weight 185g (max.) Battery packs lose capacity over time.
VIVIX-S VW Service Manual 2.6 SCU Lite (FXRP-02A) SCU Lite is a device that transmits images by wired connection between the detector and workstation. 2.6.1 Specifications Item Specifications Model FXRP-02A Power Supply Input: DC +24V 1.0A Max. Cable Connection Port 2.6.2 Rev.1.0 Gigabit Ethernet port (1ea) PoE (Power over Ethernet) Port (1ea) Dimensions (H × W × D) 108㎜ × 109㎜ × 29.
VIVIX-S VW Service Manual 2.6.3 Functions Front Side No. Name Description Indicates the status of PoE port. 1 LINK LED Green : 1 Gbps Yellow : 100 Mbps PoE (Power over Ethernet) port (1000BASE-T) 2 POE Port Communication between the detector and SCU Lite. Supplies power to the detector. Rear Side No. Name Rev.1.0 1 LAN port 2 DC power input port 3 Power Switch Description Gigabit Ethernet port (1000BASE-T) Communication between the workstation and SCU Lite.
VIVIX-S VW Service Manual 2.7 SCU mini (FXRS-04A) The SCU mini is located between the detector, the workstation and the X-ray generator to synchronize images and signals. The SCU mini has a wired or wireless connection to the detector and a wired connection to the workstation. It can also be used by wired connection with the X-ray generator (DR Trigger). 2.7.1 2.7.2 Rev.1.0 Specifications Item Specifications Model FXRP-04A Power Supply Input: DC +24V, 2A Max.
VIVIX-S VW Service Manual 2.7.3 Functions Front Side No. Name Description 1 Power switch Power switch of SCU mini. (ON / OFF) 2 Antenna Assists communications between the detector and SCU mini. Indicates status of SCU mini operation and connection. 3 Rev.1.0 Status LED Blinking green: Booting Green: Completed to boot up Blue: The detector is connected and ready to communicate.
VIVIX-S VW Service Manual Rear Side No. Name Description 1 X-ray generator interface connector (D-SUB 15pin, Female) 2 EXT_INF LAN port (Port 1, 2, 3) Gigabit Ethernet port (1000BASE-T) Port 1: Communication between the workstation and SCU mini. Port 2, 3: For configuring multiple detectors. Indicats the status of PoE port. 3 PoE status lamp Green: 1 Gbps Orange: 100 Mbps PoE (Power over Ethernet) port (1000BASE-T) 4 PoE port Communication between the detector and SCU mini.
VIVIX-S VW Service Manual 2.8 Others 2.8.1 X-ray Generator (Recommended Exposure Condition) 2.8.2 Item Recommended condition X-ray energy range 40kVp ~ 150kVp Reliability (Lifetime Dose) 74Gy or more (35 uGy x 365days x 24hrs. x 60min. x 60sec. / 15sec.) Recommended Specifications of Workstation (PC) Item Recommended specification VIVIX Setup Microsoft Windows 7 SP1 (32bit / 64bit) Operating System Microsoft Windows 8 / 8.1 / 8.
VIVIX-S VW Service Manual 3. System Configuration This section gives information about the various connection / configuration ways among the detector, SCU, workstaton and X-ray generator. You can figure out the most suitable way of system configuration through this instruction. Detector Connection Methods Diagram System Configuration Generator Interface Configuring DR Trigger Interface Rev.1.
VIVIX-S VW Service Manual 3.1 Detector Connection Methods You can organize wired or wireless connection between the VIVIX-S VW detector and SCU. You can also choose other connection ways suitable for the use environment in case of need. 3.1.1 Wireless Connection The VIVIX-S VW Detector and PC (Workstation) connect wirelessly to transfer video and data to a PC. Using the SCU mini (FXRS-04A), you can use the DR Trigger as well as AED (Auto Exposure Detection) with the generator interface.
VIVIX-S VW Service Manual Be sure to operate the wireless commnunication between the detector and SCU within a maximum of 8m. Use of multiple WLAN devices within the same frequency band may interfere with each wireless communication and cause a decline in transmission speed. Do not cover or block the wireless module of the detector. Also, do not put any shielding materials between the detector and SCU. Otherwise, the transmission speed or operable distance may be reduced.
VIVIX-S VW Service Manual We recommend you the wired connection in the following cases. If the detector needs a contant power supply. When the detector is fixed to a bucky stand or a table. If you want data communication faster than wireless connection. In a wired connection, wireless communication is disabled. Nous vous recommandons la connexion filaire dans les cas suivants. Si le détecteur nécessite une alimentation continue.
VIVIX-S VW Service Manual 3.2 Diagram 3.2.1 Block Diagram 3.2.2 Wiring Diagram Rev.1.
VIVIX-S VW Service Manual 3.3 System Configuration This section gives information about the configuration modes of VIVIX-S VW detectors, SCU, workstation and the generator. 3.3.1 SCU AP Mode In SCU AP mode, the detector and the SCU mini (FXRS-04A) are connected wirelessly, and the SCU mini and PC are connected by wired. SCU mini & Detector SCU and the detector are connected wirelessly. In this case, SCU operates as AP (Access Point) and detector as STATION.
VIVIX-S VW Service Manual If you use the detector's AED mode as the generator interface, it is unnecessary to wire the SCU and the generator. Si vous utilisez le mode DAE (détection d'exposition automatique) du détecteur comme interface de générateur, il est inutile de câbler le SCU et le générateur. 3.3.2 Tether Interface Mode Tether interface mode is the wired connection method that connecting the detector with SCU as the center.
VIVIX-S VW Service Manual 3.3.3 External AP Mode External AP mode is the wireless connection method that adding the external AP device instead of SCU. External AP & SCU The external AP and SCU are connected with a LAN cable. External AP & Detector The external AP and VIVIX-S VW detectors are connected wirelessly. In this case, the external AP and the detector operate as AP (Access Point) and STATION respectively. Other connection ways of the tether interface mode are same as the AP mode.
VIVIX-S VW Service Manual 3.3.4 AP Switching Using this function, you can conveniently change the external AP when the detector is STATION. Ex.) If you want to use one detector in both ROOM #1 and ROOM #2. 1 When the detector is in STATION, you can change the external AP by pressing the AP button for about 3 seconds. That is, the detector changes to the same SSID and key as those of the external AP to be changed.
VIVIX-S VW Service Manual 3.3.5 Sync Wireless Settings You can easily switch from tether interface mode to SCU AP mode. 1 In the tether interface mode, press the AP button for about 3 seconds to change the detector to the same SSID and KEY as the wired SCU. 2 During synchronization, the OLED display shows ‘Changing Settings’. After switching, you can check the changed settings. For details on the synchronization of wireless setup, refer to the VIVIX Setup Operation Manual.
VIVIX-S VW Service Manual 3.4 Generator Interface VIVIX-S VW detectors provide AED interface and DR Trigger interface as a generator interface method to acquire images by detecting X-ray. Mode AED Description The detector detects X-ray exposure from the generator automatically and then performs image acquisition without any cable connection. The detector and generator receive and send their signal to each other for image acquisition.
VIVIX-S VW Service Manual N ms, which is the time that the actual X-ray is to be exposed, should be shorter than the exposure time set in the detector. Otherwise, abnormal images may be obtained. N ms, which is the time that the actual X-ray is to be exposed, should be 3ms or more and 5000ms or less. Otherwise, the image may not be obtained. Make sure to follow the operating environmental condition (Temp: 0℃ ~ +40℃). Do not give impact to the product.
VIVIX-S VW Service Manual 3.4.2 DR Trigger Interface Connect the detector and X-ray generator with a generator interface cable, and then acquire images by sending and receiving their signal. Signal Processing Steps 1 The detector receives EXP_REQ signal from the X-ray generator. 2 After the detector completes to prepare image acquisition, it sends EXP_OK signal to the X-ray generator. 3 The X-ray generator confirms EXP_OK signal and generates X-rays.
VIVIX-S VW Service Manual EXP_OK est un signal de prise de vue prêt envoyé par le détecteur au générateur de rayons X. Le temps d’exposition aux rayons X défini par défaut dans le détecteur est de 500 ms et peut être modifié à l’aide des paramètres de l’utilisateur. N ms, which is the time that the actual X-ray is to be exposed, should be shorter than the exposure time set in the detector. Otherwise, abnormal images may be obtained. EXP_REQ should be 3ms or more.
VIVIX-S VW Service Manual 3.5 Configuring DR Trigger Interface To configure the DR Trigger interface, SCU mini (FXRS-04A) and the generator should be connected with a generator interface cable. Connect the one end of generator interface cable to EXT_INF port of SCU, and then connect the other end to the generator. The engineer who understands about the X-ray generator device and interface technology well should be in charge of the connection between the detector and generator interface.
VIVIX-S VW Service Manual Multipe detectors share one generator signal together. Connect a generator interface cable to one of the pin groups. The generator tramsmits and receives signal with the one selected detector for taking images. Les multi-détecteurs partagent un signal de générateur ensemble. Connectez un câble d'interface de générateur à l'un des groupes de broches. Le générateur émet et reçoit un signal avec le détecteur sélectionné pour la prise d'images. 3.5.
VIVIX-S VW Service Manual 3.5.4 EXT_INF Port Pin Map EXT_INF port is D-SUB 15 pin female connectors of SCU mini (FXRS-04A), and used for the generator interface. Product EXT_INF port SCU mini (FXRS-04A) EXT_INF1 port pin map definition (1 ~ 15) No. Signal name I/O Type Color Ch.
VIVIX-S VW Service Manual Informations de type TTL ON: VCC OFF: GND Courant: 5 ㎃ ~ 10 Tension: 12 V ~ 24 V When you plan to add interface using a user-defined pin, contact to the person in charge of Vieworks. Lorsque vous envisagez d'ajouter une interface à l'aide d'une épingle définie par l'utilisateur, contactez le responsable de Vieworks. 3.5.5 Input / Output Circuit Exposure Request Input Circuit (EXP_REQ) Exposure Respond Output Circuit (EXP_OK) Rev.1.
VIVIX-S VW Service Manual 4. Settings This section gives information about the way of installation, setting, and diagnosis to use a detector with SCU. Product Installation Device Setting Diagnosis of Devices Rev.1.
VIVIX-S VW Service Manual 4.1 Product Installation 4.1.1 This equipment should be installed by a service personnel authorized by Vieworks. This equipment must only be connected to the power with protective earth. Cet équipement doit être installé par un personnel de service autorisé par Vieworks. Cet équipement ne doit être connecté qu’à une prise de terre de protection.
VIVIX-S VW Service Manual If you use the AED interface, a generator interface cable is not needed as the detector operates by detecting X-ray automatically. Si vous utilisez l'interface AED, un câble d'interface pour générateur n'est pas nécessaire car le détecteur fonctionne en détectant automatiquement les rayons X. 2 Connect one end of the LAN cable to one of the LAN ports of SCU mini, and the other to the LAN Card Connector of workstation assigned for data transfer.
VIVIX-S VW Service Manual When the status LED is blinking green, it means power is applied and SCU is booting up. When the status LED is lit green, it means the SCU boot is complete. Lorsque le voyant d'état clignote en vert, cela signifie que l'alimentation est appliquée et que le SCU est en cours de démarrage. Lorsque le voyant d'état est allumé en vert, cela signifie que le démarrage du SCU est terminé. 4.1.
VIVIX-S VW Service Manual 4.1.4 Booting Up SCU Lite (FXRP-02A) 1 Turn on the power switch at rear side of SCU Lite (ON). 2 If you connect a detector to SCU Lite by wire, check if the LINK LED of SCU Lite lights up green. When the LINK LED lights up green, the detector and SCU are wired at 1Gbps. When the LINK LED lights up yellow, the detector and SCU are wired at 100 Mbps. Lorsque la DEL LINK est allumée en vert, le détecteur et le SCU sont câblés à 1 Gbps.
VIVIX-S VW Service Manual 4.2 Device Setting 4.2.1 Software Installation 1 After connecting all devices, prepare the following softwares to set, calibrate and operate the detector / SCU. Software Description VIVIX Device Driver (VDD) Image filter driver for acquiring images from a detector. VIVIX Setup A program for setting and managing the detector / SCU. 2 Install VIVIX Device Driver and VIVIX Setup in sequence.
VIVIX-S VW Service Manual 4.3 Diagnosis of Devices Execute the VIVIX Setup program to check if threre is any problem to operate the detector / SCU after installing and setting devices. Diagnosis Items Items Description Image Diagnoses the acquired images. Battery Pack Diagnoses the condition of a battery pack. Wired/Wireless Communication Status Diagnoses the status of wired/wireless communication. Wired/Wireless Communication Speed Diagnoses the speed of wired/wireless communication.
VIVIX-S VW Service Manual If any problem is found on the image, check if it is caused by the surrounding environment and calibrate the detector again. Contact the person in charge of Vieworks if the problem is caused by the performance of a detector. Si un problème est détecté sur l'image, vérifiez s'il est causé par l'environnement et étalonnez à nouveau le détecteur. Contactez le responsable de Vieworks si le problème est causé par les performances d'un détecteur. 4.3.
VIVIX-S VW Service Manual Since a battery pack is consumables which performance will be decreased as time passes, make sure to check its life when you use it. If a battery pack has any problems, consult service personnel in Vieworks. Parce qu’une batterie est un consommable dont les performances diminuent avec le temps, assurez-vous de vérifier sa durée de vie lors de son utilisation. Si une batterie rencontre des problèmes, consultez le personnel de maintenance de Vieworks.
VIVIX-S VW Service Manual Link speed: 100Mbps Operating Status Changing Changing communication settings Settings 2 On the initial display of the OLED display, pressing the power button or AP button for about 1 second will switch the OLED display to the second screen. On the second screen, check the SSID and KEY important for wireless connection.
VIVIX-S VW Service Manual Wireless Communication Diagnosis 1 Connect the detector wirelessly. 2 Execute VIVIX Setup and go to the Maintenance tab of the Maintenance window. 3 Check the status of wireless communication from the Network item. You can also check the communication status from VXvue or VIVIX SDK. Refer to VIVIX Setup Operation Manual for the detailed information about the wireless communication diagnosis.
VIVIX-S VW Service Manual Assurez-vous de vérifier la communication sans fil environnante pour éviter les interférences de communication. Si le module de communication sans fil du détecteur rencontre des problèmes, contactez le technicien de maintenance de Vieworks 4.3.4 Diagnosing Communication Speed 1 Execute VIVIX Setup and go to the Maintenance tab of the Maintenance dialog. 2 Click on the Download button in Image Transmission Time and check image transmission speed of the detector.
VIVIX-S VW Service Manual 4.3.5 Self-Diagnosis 1 Execute VIVIX Setup and go to the Self Diagnosis tab in the Diagnosis dialog. 2 Check the desired items to diagnose from the Test Case list. 3 Click on the Execute Test button to perform self-diagnosis. 4 Check the status and result of diagnosis for each item in the Result dialog. Click the Save to File button to save the result of a diagnosis as a file and contact the service engineer if any problem is found.
VIVIX-S VW Service Manual Self-diagnosis items of detector and measures Voltage Item Form Expected problem Measures Replace the battery pack. SOURCE Decision Problem with input power Replace the tether interface cable. Replace AC-DC adapter. MAIN Decision Problem with system power Contact a service engineer. Expected problem Measures The battery is not attached. Check if a battery is attached or not. Battery Item Detection Form Decision A defective ciruit is connected to a battery pack.
VIVIX-S VW Service Manual Memory Item Detection Form Decision File system Decision Status Information Expected problem Measures Impossible to save backup images. Contact a service engineer. Impossible to save logs. Contact a service engineer. The calibration data is inapplicable. Contact a service engineer. Impossible to save backup images. Contact a service engineer. Impossible to save logs. Contact a service engineer.
VIVIX-S VW Service Manual 5. Maintenance This section gives information about maintenance of the product. Product Initialization Detector Power Save Function How to Replace the Battery Pack Rev.1.
VIVIX-S VW Service Manual 5.1 Product Initialization If connection status of the system is not stable or setting value is not correct, you can initialize the products. 5.1.1 SCU Initialization 1 Click on the Configuration button of SCU after running the VIVIX Setup program. 2 Click on the Factory Reset button in the Configuration dialogue. 3 Wait for SCU to be initialized and rebooted automatically. 4 Make sure the SCU is initialized.
VIVIX-S VW Service Manual 5.1.2 Detector Initialization 1 Click the Maintenance button after executing the VIVIX Setup program. 2 Go to Maintenance window > Maintenance tab and click the Factory Reset button. 3 Wait for the detector until it is rebooted automatically. 4 Click the Configuration button and check if detector initialization is complete. Default value of detector initialization Item Default Value Network IP Address 169.254.1.10 Subnet Mask 255.255.0.0 Gateway 169.254.2.
VIVIX-S VW Service Manual 5.1.3 Wireless Initialization of Detector 1 Connect the detector wirelessly without connecting the tether interface cable. 2 Press and hold both the power button and AP button over 3 seconds. 3 Initialization will be conducted when the green LED is flickering. After that, the detector will reboot automatically. 4 When booting is complete, make sure the detector is initialized. Default value of wireless initialization Item Default Value Network IP Address 169.254.1.
VIVIX-S VW Service Manual 5.2 Detector Power Save Function You can extend the battery life by using the power saving function of the detector. This function works only when the battery is supplied power to the system. In other words, the power saving function does not work when the power is supplied from the SCU through the tether interface cable or from the AC-DC adapter. Cette fonction ne fonctionne que lorsque la batterie est alimentée par le système.
VIVIX-S VW Service Manual Checking Indication in Power Save Mode Item Description Normal The power indicator LED is on and the OLED display shows the initial screen. Sleep The power indicator LED is on and the OLED display shows a sleep mode icon. You can check the sleep status in VIVIX Setup or VXvue. VIVIX SDK notices sleep state. Shut Down Power indicator LED and OLED display are off.
VIVIX-S VW Service Manual 5.3 How to Replace the Battery Pack The battery pack is built into the detector. FXRD-3643VAW / FXRD-3643VAW PLUS / FXRD-4343VAW / FXRD-4343VAW PLUS can install up to two battery packs, and FXRD-2530VAW / FXRD-2530VAW PLUS can install only one battery pack. The battery pack must be replaced by a service engineer licensed or certified by Vieworks. La batterie doit être remplacée par un technicien de maintenance agréé ou certifié par Vieworks.
VIVIX-S VW Service Manual 5.3.1 Replacing the Battery of FXRD-3643VAW / FXRD-3643VAW PLUS / FXRD-4343VAW / FXRD-4343VAW PLUS How to remove the battery pack 1 Remove the deco sheet attached to the battery pack cover. 2 Loosen the bolts that secure the battery pack cover using a hexagon wrench, and then open the battery pack cover. 3 Take the battery pack out of the detector. 4 Disconnect the cable connected to the battery pack. Rev.1.
VIVIX-S VW Service Manual Before removing the battery pack from the detector, turn off the detector. If you press and hold the power button for about 3 seconds, the power indicator LED and the OLED display turn off, then the system is powered off. Disconnect wired power before removing the battery pack from the detector. If a tether interface cable or AC-DC adapter is attached, remove it from the detector. Avant de retirer la batterie du détecteur, éteignez le détecteur.
VIVIX-S VW Service Manual 5.3.2 Replacing the Battery of FXRD-2530VAW / FXRD-2530VAW PLUS 1 Remove the deco sheet attached to the battery pack cover. 2 Loosen the bolts that secure the battery pack cover using a hexagon wrench, and then open the battery pack cover. 3 Take the battery pack out of the detector. 4 Disconnect the cable connected to the battery pack. How to attach the battery pack 1 Connect the battery pack to the cable. 2 Put the battery pack into the detector.
VIVIX-S VW Service Manual 6. Troubleshooting Rev.1.
VIVIX-S VW Service Manual 6.1 Troubleshooting 6.1.1 Troubleshooting Guide When you encounter problems while using the equipment, search for the table below for the problem or error messages and try the solutions. If the problem persists, turn off the detector and consult your sales representative or a distributor. Please refer to the details of the following symptoms or error messages. Troubleshooting must be performed by service engineer who is authorized by Vieworks.
VIVIX-S VW Service Manual 1 Check the cable connections (AC-DC adapter, power cable, LAN cable) of the SCU. 2 Turn on and off SCU. Solutions 3 Replace the power cable. 4 Replace the AC-DC adapter. 5 Replace the LAN cable. 6 Replace SCU. 6.1.4 Communication Test is failed Category Description Symptom The communication connection fails or is lost. A transmission error occurs and abnormal images or data are sent.
VIVIX-S VW Service Manual 2 In DR trigger mode, check the generator interface connection. 3 Replace the generator interface cable. 4 Replace the SCU. 5 Replace the detector. 6.1.6 Getting Abnormal Images Category Description Symptom The image sent to the PC is abnormal. Problems with calibration data Expected Causes Problems with setting exposure time Wrong operating environment Detector failure. 1 Perform the calibration again.
VIVIX-S VW Service Manual 7. Information This section gives overview information for service and warranty of the product. Service Information Warranty Revision History Rev.1.
VIVIX-S VW Service Manual 7.1 Service Information 7.1.1 Product Lifetime The estimated product lifetime may be up to seven (7) years under the appropriate regular inspection and maintenance. 7.1.2 Regular Inspection and Maintenance In order to ensure the safety of patients, operating personnel and third parties, and to maintain the performance and reliability of the equipment, be sure to perform regular inspection at least once a year.
VIVIX-S VW Service Manual 7.2 Warranty Vieworks Co., Ltd. follows the contract with each customer regarding the product warranty period. If any such product proves defective during this warranty period, Vieworks at its option, either will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the defective product.
FCC Compliance This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. Operation is subject to the following tow conditions. This device may not cause harmful interference. This device must accept any interference received, including interference that may cause undesired operation.
VIVIX-S VW Service Manual 7.3 Rev.1.0 Revision History Version Date Descriptions 1.
VIVIX-S VW Service Manual Vieworks Co., Ltd. (Gwanyang-dong) 41-3, Burim-ro 170beon-gil, Dongan-gu, Anyang-si, Gyeonggi-do, 14055 Republic of Korea Telephone: +82-70-7011-6161 Fax: +82-31-386-8631 Homepage: http://www.vieworks.com European representative: Obelis s.a Bd. Général Wahis 53 1030 Brussels, BELGIUM Tel: +(32) 2.732.59.54 Fax: +(32) 2.732.60.03 E-mail: mail@obelis.net Rev.1.