Service Manual

Rev.1.0
Page 24 of 90 D-19-616
VIVIX-S VW Service Manual
2.3 FXRD-4343VAW, FXRD-4343VAW PLUS
2.3.1 Specifications
Item
Specifications
Model
FXRD-4343VAW
FXRD-4343VAW PLUS
Image Sensor
TFT: a-Si (Amorphous Silicon)
X-ray Scintillator Type
FXRD-4343VAW: Csl A type
FXRD-4343VAW PLUS: CsI B type
Pixel Pitch
0.14 (140)
Field of View
43cm x 43cm (17 x 17)
Active Area (H x V)
430.08 × 430.08
Active Array
3072 x 3072 pixels
Effective Area
426.72 x 426.72
Effective Array
3048 x 3048 pixels
Grayscale
16 bit
Spatial Resolution
Min. 3.5 lp/
Image Acquisition Time (Wired)
Up to 3 sec. (Set exposure time to 500ms / except exposure time)
Image Acquisition Time
(Wireless)
Up to 4.5 sec. (IEEE802.11n, MIMO 2x2, 5GHz, 40MHz)
Up to 3 sec. (IEEE802.11ac, MiMO 3x3, 5GHz, 80MHz)
(Exposure time is set to 500ms, except exposure time)
Cycle Time
Min. 4 sec. (with optimal wired / wireless environment, exposure
time is set to 500ms, excluding software processing time)
X-ray Synchronous Control
AED (Auto Exposure Detection)
DR Trigger (External Line Trigger)
Power Supply
Powered from SCU via tether interface cable: DC 24V, Max. 1.0A
Powered via AC-DC adapter: DC 18V, Max. 4.44A
Powered via 2 battery packs: DC 9 ~13.2V, Max. 78.54Wh
Power Consumption
Max. 24W (without battery charged)
Max. 80W (when charging battery)
Operating Time
(Early life of battery)
One battery pack
6.5 hours (image acquisition every 100 seconds), 8 hours (standby)
Two battery packs
13 hours (image acquisition every 100 seconds), 16 hours (standby)
Dimensions (H × W × D)
460.0 × 460.0 × 15.0
Weight (including battery pack)
With one battery pack : 3.5kg
With two battery packs : 3.7kg
Image Transmission
Wired: Gigabit Ethernet (1000BASE-T) via Power over Ethernet (PoE)
Wireless: IEEE 802.11n / ac (2.4GHz / 5GHz), 3 antennas
Data Transmission Speed (Wired)
Max. 1Gbps
Data Transmission Speed
(Wireless)
Max. 450Mbps (IEEE802.11n, MIMO 2x2, 5GHz, 40MHz)
Max. 1300Mbps (IEEE802.11ac, MIMO 3x3, 5GHz, 80MHz)