Service Manual

Rev.1.0
Page 28 of 90 D-19-616
VIVIX-S VW Service Manual
2.4 FXRD-2530VAW, FXRD-2530VAW PLUS
2.4.1 Specifications
Item
Specifications
Model
FXRD-2530VAW / FXRD-2530VAW PLUS
Image Sensor
TFT: a-Si (Amorphous Silicon)
X-ray Scintillator Type
FXRD-2530VAW: Csl A type
FXRD-2530VAW PLUS: CsI B type
Pixel Pitch
0.124 (124)
Field of View
25cm x 32cm (10 x 12)
Active Area (H x V)
253.95 × 317.44
Active Array
2048 x 2560 pixels
Effective Area
251.0 x 314.5
Effective Array
2024 x 2536 pixels
Grayscale
16 bit
Spatial Resolution
Min. 4.0 lp/
Image Acquisition Time (Wired)
Up to 3 sec. (Set exposure time to 500ms / except exposure time)
Image Acquisition Time
(Wireless)
Up to 4.5 sec. (IEEE802.11n, MIMO 3x3, 5GHz, 40MHz)
Up to 3 sec. (IEEE802.11ac, MiMO 3x3, 5GHz, 80MHz)
(Exposure time is set to 500ms, except exposure time)
Cycle Time
Min. 4 sec. (with optimal wired / wireless environment, exposure
time is set to 500ms, excluding software processing time)
X-ray Synchronous Control
AED (Auto Exposure Detection)
DR Trigger (External Line Trigger)
Power Supply
Powered from SCU via tether interface cable: DC 24V, Max. 0.625A
Powered via AC-DC adapter: DC 18V, Max. 2.78A
Powered via 2 battery packs: DC 9 ~13.2V, Max. 39.27Wh
Power Consumption
Max. 15W (without battery charged)
Max. 50W (when charging battery)
Operating Time
(Early life of battery)
One battery pack
7 hours (image acquisition every 100 seconds), 8 hours (standby)
Dimensions (H × W × D)
287.0 × 350.0 × 15.0
Weight (including battery pack)
1.9kg
Image Transmission
Wired: Gigabit Ethernet (1000BASE-T) via Power over Ethernet (PoE)
Wireless: IEEE 802.11n / ac (2.4GHz / 5GHz), 3 antennas
Data Transmission Speed (Wired)
Max. 1Gbps
Data Transmission Speed
(Wireless)
Max. 450Mbps (IEEE802.11n, MIMO 3x3, 5GHz, 40MHz)
Max. 1300Mbps (IEEE802.11ac, MIMO 3x3, 5GHz, 80MHz)
The longer you use the battery, the smaller its capacity.
Plus vous utilisez la batterie, plus sa capacité est réduite.