4.1.4 Booting Up SCU Lite (FXRP-02A)
1
Turn on the power switch at rear side of SCU Lite (ON).
2
If you connect a detector to SCU Lite by wire, check if the LINK LED of SCU Lite lights up green.
When the LINK LED lights up green, the detector and SCU are wired at 1Gbps.
When the LINK LED lights up yellow, the detector and SCU are wired at 100 Mbps.
Lorsque la DEL LINK est allumée en vert, le détecteur et le SCU sont câblés à 1 Gbps.
Lorsque la DEL LINK est allumée en jaune, le détecteur et le SCU sont câblés à 100 Mbps.
4.1.5 Booting up the Detector
1
Press the power button on the detector for about 1 second.
2
Check that the detector's power LED is lit green.
3
Check the battery status (remaining, charged) and communication connection status (wired / wireless,
detector AP / STATION, wireless connection strength, etc.) on the detector's OLED display.
If the power LED flashes green, power is applied, and the detector is booting up.
When the power indicator LED is lit green, the detector has finished booting.
If you press and hold the power button for about 3 seconds, the power indicator LED and
the OLED display turn off, then the system is powered off. However, the battery pack can
be charged when the SCU is powered via a tether interface cable or an AC-DC adapter is
attached.
Si le voyant d'alimentation clignote en vert, l'alimentation est appliquée et le détecteur
démarre.
Lorsque le voyant d'alimentation est allumé en vert, le détecteur a fini de démarrer.
Si vous maintenez le bouton d'alimentation enfoncé pendant environ 3 secondes, le
voyant d'alimentation et l'écran OLED s'éteignent, puis le système est mis hors tension.
Toutefois, la batterie peut être chargée lorsque le SCU est alimenté via un câble
d'interface d'attache ou lorsqu'un adaptateur CA-CC est connecté.