6.1 Troubleshooting
6.1.1 Troubleshooting Guide
When you encounter problems while using the equipment, search for the table below for the problem or
error messages and try the solutions. If the problem persists, turn off the detector and consult your sales
representative or a distributor. Please refer to the details of the following symptoms or error messages.
Troubleshooting must be performed by service engineer who is authorized by Vieworks. If
an unqualified person performs troubleshooting on the system resulting in damaging the
detector, software or hardware, then the Vieworks or its representative is not responsible
for the detector repair regardless of remain warranty. For more detailed information, refer
to <7.1 Service Information> and <7.2 Warranty>.
Le dépannage doit être effectué par un technicien de maintenance agréé par Vieworks. Si
une personne non qualifiée effectue un dépannage sur le système entraînant une
détérioration du détecteur, du logiciel ou du matériel, Vieworks ou son représentant n'est
pas responsable de la réparation du détecteur, quelle que soit la garantie restante. Pour
plus d'informations, reportez-vous aux sections <7.1 Informations sur la maintenance> et
<7.2 Garantie>.
6.1.2 Fail to Turn the Detector On
Category
Description
Symptom
Failed to turn the power of the detector.
Expected Causes
The battery pack is discharged.
The battery pack is attached wrongly.
The battery pack is broken.
The detector is broken.
Solutions
1 Charge a battery pack
2 Attach a battery pack again.
3 Replace with another battery pack.
4 Replace with another detector.
6.1.3 The Power Switch of SCU or Status LED is not worked normally
Category
Description
Symptom
The power switch of SCU is not working.
The status LED of SCU is not responding.
Expected Causes
Problems with the connection of power cable.
Power cable is broken.
AC-DC adapter is broken.
LAN cable is broken.
SCU is broken.