Service Manual

Rev.1.0
Page 84 of 90 D-19-616
VIVIX-S VW Service Manual
Solutions
1 Check the cable connections (AC-DC adapter, power cable, LAN cable) of the SCU.
2 Turn on and off SCU.
3 Replace the power cable.
4 Replace the AC-DC adapter.
5 Replace the LAN cable.
6 Replace SCU.
6.1.4 Communication Test is failed
Category
Description
Symptom
The communication connection fails or is lost.
A transmission error occurs and abnormal images or data are sent.
Expected Causes
Network connection problem
Network setting problem
PC environment setting problem
Wireless environment environment problem
SCU (or external AP) failure
Detector failure
Solutions
1 Check the connection of network cable between workstation and SCU (or external
AP).
2 Check if the accurate network cable is used or not. (CAT 5E or 6)
3 Set the network information of workstation, SCU (or external AP) and detector
again.
4 Set the workstation environment again such as firewall setting.
5 Check surrounding wireless communication environment. If there is an environment
that prevents wireless communication, remove the obstacle or change the wireless
communication settings to prevent the interference.
6 Reboot the detector and SCU.
7 Replace SCU (or external AP).
8 Replace the detector.
6.1.5 There is no problem in communication connection, but no image is acquired
Category
Description
Symptom
The image cannot be sent to PC even though the X-ray is exposed.
Expected Causes
Problems with the generator interface
Problems with the generator interface connection
Generator interface cable failure
SCU failure
Detector failure
Solutions
1 Check the generator interface setting of the detector. (AED mode, DR Trigger
mode)