2 In DR trigger mode, check the generator interface connection.
3 Replace the generator interface cable.
4 Replace the SCU.
5 Replace the detector.
6.1.6 Getting Abnormal Images
Category
Description
Symptom
The image sent to the PC is abnormal.
Expected Causes
Problems with calibration data
Problems with setting exposure time
Wrong operating environment
Detector failure.
Solutions
1 Perform the calibration again.
2 Make sure that the detector is set to be applied calibaration data.
3 Set the exposure time to be longer than the X-ray exposure time.
4 Check the operating environment (temperature, humidity, shock, vibration, X-ray
energy range).
5 Replace the detector.
6.1.7 Battery Pack or Installation Part of Battery is Getting Hot
Category
Description
Symptom
The fully charged battery pack runs down quickly.
Expected Causes
Reduced performance (capacity) with prolonged use of the battery pack
Using the battery pack in low or high temperature environments
Solutions
1 If the battery pack is to be used for a long time, replace it with a new battery
pack. (The battery pack is a consumable.)
2 Use the battery pack at room temperature. In cold or hot environments, the
battery pack's capacity decreases.