Contents Introduction ............................................................ 5 Safety Information ................................................. 6 General ...................................................................................6 Safe handling of batteries................................................7 Service & Repairs .................................................................8 About this device .................................................... 9 Proper use...................
Setting the date & time ........................................ 22 Putting on the wrist cuff ...................................... 24 Correct position for measurement ...................... 25 Automatic measurement .............................................. 28 Arrhythmia detection ................................................... 28 User settings ......................................................... 29 Selecting user....................................................................
Introduction Thank you for purchasing the Welby® Wrist Blood Pressure Monitor. This fully automatic device measures blood pressure and pulse rate and checks detects irregular heart beat fast and easily and then displays the results on the LCD screen. It works using the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading.
Safety Information General • This device is intended for home use only. • Do not use this device for self-diagnosis or self-treatment. Contact your physician or other licensed health care provider for specific information about your blood pressure. • Keep plastic packaging out of the reach of children. There is a risk of suffocation. • Keep the device and its accessories out of reach of children. • Do not inflate (with the button) until the device is attached correctly to the wrist.
Safe handling of batteries Batteries may contain combustible materials. If not handled correctly, batteries can leak, become very hot, ignite or even explode, which could damage your device and/or your health. It is very important that you follow the instructions below: • Keep batteries away from children. If batteries are mistakenly swallowed, call poison control or 911 immediately.
inserted to avoid short circuits. • Remove discharged batteries from the device immediately. • Remove the batteries from the device if you do not intend to use it for an extended period. • Replace all of the old batteries in device with new batteries of the same type. Service & Repairs Please contact the the Customer Care line if you have questions about the device or experience technical problems. Never try to open or repair the device yourself.
About this device The device is used to measure blood pressure at the wrist and can save 60 measurement values in each of two memory spaces. Proper use This device is intended for use in dry, indoor areas only. The device is only intended for private use and not for clinical or commercial use. Using the device under extreme environmental conditions may damage it. 9 13400 EN Aldi US RC6 Content.indd 7 20.04.
Frequently Asked Questions What is blood pressure? Blood pressure is the force exerted by blood against the walls of the aterties as blood is pumped through the circulatory system and is measured in millimeters of mercury (mmHg). The higher the pressure, the harder the heart has to work to supply blood to the vital organs in the body.
What is High Blood Pressure and how is it controlled? High Blood Pressure, also called by its medical term Hypertension, is an abnormally high arterial blood pressure. If left unattended, it can cause many health problems including stroke and heart attack.
How does the blood pressure monitor work? Your blood pressure monitor contains high quality components for measuring and to display the results on the display using the oscillometric method of blood pressure measurement. It also measures and displays your pulse. The device measures blood pressure indirectly using a cuff around the wrist. Inflating the cuff presses the arteries together so that no more blood can flow through them.
Measuring range The blood pressure monitor shows your blood pressure (SYS. = systolic and DIA. = diastolic value), pulse and the time and date on the display. Blood pressure is measured in mmHg (millimetres of mercury). This devices classifies blood pressure measurements based on the definition and standards set by the World Health Organisation (WHO). WHO defines high blood pressure as any measurements above 140 mmHg (systolic) and 90 mmHg (diastolic).
Blood pressure classification according to WHO in mmHg 1 Normal range 2 Borderline high 3 High blood pressure range 14 13400 EN Aldi US RC6 Content.indd 12 20.04.
Range Low blood pressure Normal blood pressure Borderline high blood pressure Mild high blood pressure Moderate high blood pressure Severe high blood pressure Systolic mmHg Diastolic mmHg Indicator Under 90 Under 50 Green 91–129 51–84 Green 130–139 85–89 Green 140–159 90–99 Yellow 160–179 100–109 Orange Over 180 Over 110 Red 15 13400 EN Aldi US RC6 Content.indd 13 20.04.
About your measurement results If the measurements obtained give you cause to be concerned about your health, consult with your physician or licensed health care professional. Do not change your current treatments and/or dosages of medication(s) as a result of your measurements unless you have discussed this first with your doctor! By using your device regularly and recording your results, you can get a clear overview of your blood pressure. Everybody’s blood pressure changes constantly.
Before taking your blood pressure measurement Note the following points so that the measurement results give a realistic representation of your blood pressure: • Always measure your blood pressure at the same time of the day. • Measure your blood pressure when you are relaxed. Try to rest for fifteen minutes before measuring your blood pressure. • You should not eat, drink, smoke or exercise for at least 60 minutes before taking a measurement.
• Please ensure that you always measure your blood pressure under the same conditions! Also note the information about maintaining the correct posture during measurement on page 25. Different measurement conditions can lead to markedly different results. If you have a result that differs greatly from your normal blood pressure measurement, you should take measurement again after a short period of time in order to exclude the factors listed above that can lead to an incorrect measurement.
Initial operation Device overview 1 User button 2 Clock button 3 ON/OFF button 4 Battery compartment 5 Cuff 6 Save button 7 WHO color indicator 8 Display 19 13400 EN Aldi US RC6 Content.indd 17 20.04.
Display 1 AVG: Average blood pressure 2 Battery indicator 3 M: Display of stored measurement results 4 Systolic blood pressure 5 Diastolic blood pressure 6 Pulse measurement 7 Pulse display with arrhythmia detection indicator 8 User selection 9 Blood pressure range indicator 10 Cuff inflation symbol 11 Date and time 20 13400 EN Aldi US RC6 Content.indd 18 20.04.
Inserting batteries Remove the cover of the battery compartment at the notch in the direction of the arrow. Insert two 1.5 V R03/LR03/ AAA alkaline batteries. Replace the cover of the battery compartment. It is time to replace the batteries: • when the battery symbol on the display is only half shown and the error message E6 is displayed, or • when the display remains blank after pressing the button. 21 13400 EN Aldi US RC6 Content.indd 19 20.04.
Always change both batteries! Never use new and old batteries together or different types of batteries. Do not use rechargeable batteries for operating this device. Setting the date & time After you have inserted the batteries, the device will change into settings mode. The year flashes on the display. Press the M button until the correct year is displayed. Confirm your selection by pressing the clock button . The display will then change to the month setting.
Confirm your selection by pressing the clock button and the display will change to the hour setting. Press the M button until the current time is displayed. Confirm your selection by pressing the clock button and the display will change to the minute setting. Press the M button until the correct time is displayed. Confirm your selection by pressing the clock button and now the current time and date will appear on the display. Switch off the device by pressing the ON/OFF button .
Putting on the wrist cuff Remove clothing from left arm or roll up sleeve making sure it is not too tight on your arm. Attach the wrist cuff in such a way that it is approx. 1/2” away from the palm of your hand. Wrap the cuff tightly around your wrist so that there is no gap between the cuff and your wrist. Do not put over clothing. Wrap the cuff according to the shape of your wrist. If the cuff is not firmly attached, the pressure will increase and the wrist will become numb.
Correct position for measurement To measure your blood pressure, sit comfortably on a chair, take 5 or 6 deep breaths and relax. Place your left elbow on a table or other supporting surface so that the the wrist cuff is at heart height. You can also hold your left arm at heart height in front of your chest and support it using your right arm. 25 13400 EN Aldi US RC6 Content.indd 23 20.04.
It is important to measure your blood pressure at home and at the same time each day so that the blood pressure, which can change over the course of a day, is recorded properly. Record your measurements in the chart at the end of this instruction manual and discuss with your physician so that you can receive medical advice based on an evaluation of these measurements. Blood pressure measured at the wrist can vary from that measured at the upper arm.
The battery symbol, the blinking cuff inflation symbol and the selected user are shown on the display. Do not move the cuff, your arm or hand muscles during measurement. After a few seconds, the cuff will inflate. When the cuff pressure is stable, the device will start measuring automatically. If you want to stop the measurement, press the ON/OFF button and the cuff will deflates. If you want to repeat a measurement, you should wait at least five minutes. The cuff must be fully deflated.
Automatic measurement If the cuff pressure is too low, the device stops measuring and inflates the cuff again. Measurement starts when the correct cuff pressure has been reached. The device also measures your pulse. The value displayed will fall during measurement. When the measurement is complete, the cuff will deflate by itself and the systolic and diastolic blood pressure as well as the pulse and the pulse symbol will appear on the display.
User settings The blood pressure monitor has two memory banks, which allow two different users to save up to 60 measurements each. The device calculates the average value from the last three measurements. Selecting user Press the clock button or user button to switch the device on. The current time and date, as well as the current selected user, will appear on the display. If necessary, press the user button again to switch between the two memory banks. Press the ON/OFF button device.
Retrieving stored measurement results Press the user button to switch on the device. Press the user button again to select the desired user. Press the M button, the average measurement will be shown on the display. This is indicated on the display with AVG. Press the M button repeatedly in order to display the results of the last measurements. Press the ON/OFF button device. to switch off the 30 13400 EN Aldi US RC6 Content.indd 28 20.04.
Deleting memory values You can delete individually saved measurements or all of the measurements for each user. Press the ON/OFF button device. Press the user button desired user. Press the M button, the average measurement will be shown on the display. Select the measurement to be deleted with the M button. Press and hold down the M button for approx. 3 seconds. CL and two lines will be shown on the display. to switch on the again to select the 31 13400 EN Aldi US RC6 Content.
Press the user button , CL and 00 will be shown on the display. The value has now been deleted. In order to delete all of the saved measurements for the selected user, repeat the first three steps until the average measurement is shown on the display. This is indicated on the display by AVG. Press and hold down the M button for approximately 3 seconds. CL and two lines will be shown on the display. 32 13400 EN Aldi US RC6 Content.indd 30 20.04.
Press the user button , CL and 00 will be shown on the display. All of the saved measurements for the selected user have now been deleted. 33 13400 EN Aldi US RC6 Content.indd 31 20.04.
Caring for your blood pressure monitor Cleaning/storage Only use a dry, soft cloth to clean the device. Do not use chemical solutions or cleaning products because these may damage the device’s surface or markings. Store the device in the supplied case when not in use. Do not bend the preshaped cuff outwards. The cuff should neither be washed nor dry cleaned. If the device is not going to be used for long periods, remove the batteries. 34 13400 EN Aldi US RC6 Content.indd 32 20.04.
Waste disposal Packaging Your device has been packaged to protect it from damage in transit. The packaging is made of environmentally friendly material which can be recycled. Device At the end of its service life, the device should not be disposed of with the household trash. Please enquire about options for environmentally friendly disposal. Batteries Do not dispose of used batteries in the household trash! They should be disposed of in an environmentally appropriate way. 35 13400 EN Aldi US RC6 Content.
Troubleshooting If there are error messages or incorrect measurements displayed, remove the batteries and insert them again after about one minute. Error Possible causes Display is blank The batteries are dead. E1 E2 Solutions Replace both batteries with new 1.5 V R03/LR03/AAA batteries. The cuff is not Make sure cuff worn correctly is worn properly and take a new measurement. You moved while Sit still and take a a measurement new measurement. was being taken. 36 13400 EN Aldi US RC6 Content.
Error E3 E4 E5 E6 Possible causes solutions and The cuff is not Fit the cuff correctly attached correctly. and take a new measurement. An error has Take a new occurred in the measurement. measurement Cuff pressure Check the fitting of above 300 mmHg the cuff and take a new measurement. Batteries are too weak Replace both batteries with new 1.5 V R03/ LR03/AAA batteries. 37 13400 EN Aldi US RC6 Content.indd 35 20.04.
Technical data Measurement method: Oscillometric Measuring range Blood pressure: 30 ~ 280 mmHg Pulse: 40 ~ 180 beats/min. Power supply: 2 x 1.5 V alkaline batteries R03/LR03/AAA Display: LCD Memory spaces: 2 x 60 measurement results Inflation: Automatically by pump Air release: Automatically via a one-way valve Measuring accuracy Blood pressure: +/- 3mmHg pressure indicator Pulse: +/- 5% of value displayed 38 13400 EN Aldi US RC6 Content.indd 36 20.04.
Temperatures Operating temperature: +50 ~ +104 °F, at humidity 30–85% Storage temperature: −4 ~ +140 °F, at humidity 10–95% Cuff size: 5.3” ~ 7.68” Specifications subject to change without notice. 39 13400 EN Aldi US RC6 Content.indd 37 20.04.
FCC Statement NOTE: POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only) This product has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
circuit different from that to which the receiver is connected. • Consult the dealer or an experienced radio/TV technician for help. You may also find helpful the following booklet, prepared by the FCC: “How to Identify and Resolve Radio-TV Interference Problems.” This booklet is available from the U.S. Government Printing Office, Washington D. C. 20402.
Blood pressure reading log User 1 Date & Sys Time Dia Pulse Wt. Notes 42 13400 EN Aldi US RC6 Content.indd 40 20.04.
User 2 Date & Sys Time Dia Pulse Wt. Notes 43 13400 EN Aldi US RC6 Content.indd 41 20.04.
Limited Warranty PLEASE READ THIS LIMITED WARRANTY CAREFULLY TO UNDERSTAND YOUR RIGHTS AND OBLIGATIONS. MEDION warrants that the Welby® Premium Wrist Blood Pressure Monitor (MD 13400, Item 1990) accompanied by this limited warranty is free from manufacturing defects in material or workmanship for a period of 36 months from the date of original purchase. Please save your original cash register receipt as it may be required to validate original date of purchase.
limited warranty extends to repaired products and replacement products only through the end of the original limited warranty period. In no event shall MEDION or any of its distributors or resellers be liable for incidental, indirect or consequential damages relating to the product (including any repaired product or replacement product) or its use. Some States do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation and exclusion may not apply to you.
To obtain a RMA please contact MEDION Toll Free Customer Care at: 866-633-4660 Daily 8:00 am - 12:00 am CST except holidays Email Support (Available 24 x 7, with response in 24 hours or less) 46 13400 EN Aldi US RC6 Content.indd 44 20.04.