Datasheet

2300 System
Vishay Micro-Measurements
www.vishaymg.com micro-measurements@vishay.com Document Number: 11255
22 Revision: 18-Oct-06
Signal Conditioning Amplifier
FEATURES
Accepts all strain gage inputs (foil and piezoresistive),
potentiometers, DCDT's, etc
Selectable bridge excitation, 0.7 to 15Vdc (11 steps), plus
0.2 to 7Vdc continuously variable
Fully adjustable calibrated gain from 1 to 11 000
Dual-range (±5000µ
ε and ±25 000µε) automatic bridge
balance, with "keep-alive" power to preserve balance for
months without external power
All bridge completion built in, including 120- or 1000- and
350-ohm dummies
Dual polarity two-step double shunt calibration
Bandpass:
60kHz (-0.5dB)
140kHz (-3.0dB)
Switchable active filter - a 6-pole Butterworth is standard
Two simultaneous buffered outputs
Playback mode to filter and observe or re-record previously
recorded low-level data
Input impedance above 100 megohms
DESCRIPTION
The 2300 System conditions and amplifies low-level signals
to high-level outputs for multiple-channel, simultaneous dy-
namic recording and display on external devices.
Among its features, each 2310B Module includes a built-in
power supply, active filtering, two simultaneous outputs,
playback mode, wide frequency response, and voltage injec-
tion bridge balance.
Up to ten 2310B Modules can be mounted in a Model 2350
Rack Adapter; or up to four modules in a Model 2360B
Portable Enclosure; or, a single 2310B can serve as a stand-
alone unit using the 2310-A20 Line Cord and Stabilizer.