Datasheet
System 5000
Vishay Micro-Measurements
StrainSmart
®
Data Acquisition System
Document Number: 11269 micro-measurements@vishay.com www.vishaymg.com
Revision 25-Aug-06 37
MEASUREMENT RANGES: ±0.5, ±1, ±2.5, ±5Vrms
RESOLUTION: 15.25, 30.5, 76.2, 152.5 μVrms
CALIBRATION: Excitation sample
EXCITATION:
3.0Vrms, 5000Hz or 2500Hz sine wave.
Software-selectable
Accuracy:
±5mVrms typical
Current:
±250mA max (±50mA per channel).
Over-current protected
Load Regulation:
<0.1% of full scale for a load variation of 10% to 100% of
full load
Temperature Stability:
Better than ±0.05%/°C
FILTER:
Type:
Four-pole Butterworth
Cutoff Frequency (-3dB):
5HZ