SENSOR CARD SPECIFICATIONS
MODEL 6010A STRAIN GAGE CARD
• Supports software identification and setup of each type of
plug-in card
• Complete strain gage signal conditioner with 16-bit
analog-to-digital converter
• Programmable digital filter
• Programmable excitation supply per channel. The supply
is settable to 0, 0.5, 1, 2, 5, and 10V. Up to 50mA of
excitation current is available on each channel. Remote
sense is provided for full-bridge transducers.
• Programmable gain to complement the excitation steps of
1, 2, 5, and 10V. Full-scale input range will be ±16 383με.
An excitation setting of 0.5V will use the 1V gain range, but
with one-half the resolution. Gain settings are independent
per channel.
• Internal bridge completion resistors: 120Ω, 350Ω, and
1000Ω dummy resistors (jumper selectable); 1000 internal
half bridge
• Programmable coarse balance range of ±16 300με
(4096με steps)
• Fixed low-pass anti-aliasing filter (six-pole)
• Two programmable shunt calibration circuits
• Input connections to user's strain gage via nine-pin D-sub
connector
CHANNELS: One per card.
INPUTS:
Strain Gages:
120Ω, 350Ω,1000Ω quarter bridges; 60Ω to 5000Ω half
and full bridges.
Jumper-selectable completion resistors
(0.01% ± 2.5ppm/°C typical).
Measurement Range:
Normal range mode: ±16 380με
High range mode: ±163 800με
Low range mode: ±1638με
Resolution:
Normal range mode: 0.5με
High range mode: 5με
Low range mode: 0.05με
Strain Gage Based Transducers:
60Ω to 5000Ω impedance
Measurement Range:
Normal range mode: ±8mV/V
High range mode: ±80mV/V
Low range mode: ±0.8mV/V
Resolution:
Normal range mode: 0.25μV/V
High range mode: 2.5μV/V
Low range mode: 0.025μ
V/V
Input Impedance:
220MΩ each input.
Source Current: ±25nA max.
Input Connector: Nine-pin D-sub style.