Datasheet
System 6000
Vishay Micro-Measurements
StrainSmart
®
Data Acquisition System
Document Number: 11270 micro-measurements@vishay.com www.vishaymg.com
Revision 14-Nov-06 41
AMPLIFIER:
Zero Temperature Stability:
±1.5μV/°C RTI, ±100μV/°C RTO, after 30-minute
warm-up.
DC Gain Accuracy & Stability:
(±0.1% at 23°C) ±50ppm/°C
Common-Mode Rejection (dc to 60Hz):
100dB typical
Common-Mode Voltage:
±6V typical
AC Accuracy: (typical):
Input Frequency/Bandwidth: 500/3000 50/200
Spurious Free Dynamic Range: 100dB 110dB
Signal to Noise: 90dB 95dB
Signal to Distortion: 100dB 110dB
Coarse Balance Range:
±99% of measurement range (typically ±16 300με)
CALIBRATION:
Two shunt calibration points are available on each channel.
Switch-selectable.
Calibration switches, A and B, are software selectable.
EXCITATION:
0.0, 0.5, 1.0, 2.0, 5.0, and 10.0Vdc. Software-programmable.
Accuracy:
±3mV typical
Current:
50mA max. Over-current protected.
Load Regulation:
<0.05% of full scale for a load variation of 10% to 100% of
full load
Temperature Stability:
Better than ±0.005%/°C
Remote Sense:
15Ω maximum lead resistance.
A/D CONVERTER :
Type:
16-bit successive approximation with integrated sample
and hold.
Integral Linearity Error:
±2 LSB
FILTERS:
Linear phase, analog, 6-pole anti-aliasing filter, and 256-tap,
programmable, FIR digital filter (lowpass)
Passband Frequency:
User-selectable 1Hz to 4kHz
ANALOG OUTPUT: (6010A version only)
Type:
±5.00V max for typical full-scale input of ±16 380με
Output Load:
2000Ω min
Bandwidth:
DC to 15kHz (-0.5dB typical)
MODEL 6020A THERMOCOUPLE CARD
• Complete thermocouple signal conditioner with 16-bit
analog-to-digital converter
• Programmable digital filter
• Programmable common cold-junction reference
• Compensation is provided for J, K, T, N, E, R, S, and B
thermocouple types
• Fixed low-pass anti-aliasing filter (six-pole)
• Connections to user's thermocouple circuit via a
removable three-terminal screw connector
CHANNELS: One per card.
INPUTS:
Thermocouple types J, K, T, E, N, R, S, B. Built-in electronic
cold-junction compensation. Software-selectable
Open sensor detection
Input Impedance: 10MΩ differential, 100KΩ common
mode.
Source Current: ±0.5nA typical; ±5nA max.
INPUT CONNECTOR: Three-position screw terminal
AMPLIFIER:
Zero Temperature Stability:
±1.5μV/°C RTI, ±100μV/°C RTO, after 30-minute
warm-up.
DC Gain Accuracy & Stability:
0.05% at 23°C ± 50ppm/°C
Common-Mode Rejection (dc to 60Hz):
100dB typical
Common-ModeVoltage:
±6V typical