Datasheet

System 6000
Vishay Micro-Measurements
StrainSmart
®
Data Acquisition System
www.vishaymg.com micro-measurements@vishay.com Document Number: 11270
44 Revision: 14-Nov-06
MODEL 6050 PIEZOELECTRIC CARD
Complete piezoelectric signal conditioner with 16-bit
analog-to-digital converter
Supports both VM (voltage mode, low impedance) and CM
(charge mode, high impedance) type piezoelectric
transducers
Programmable constant current excitation supply for VM
transducers is software settable to 1, 2, 4, 5, 10 and 20mA
Programmable gain steps of 1, 2, 5, and 10 for VM
transducers and steps of 1, 2, 5, 10, 20, 50 and 100 for CM
transducers
Programmable digital filter
Fixed low-pass anti-aliasing filter (six-pole)
Input connections to user's transducer via BNC connector
CHANNELS: One per card.
INPUTS: Voltage mode (VM) or charge mode (CM)
piezoelectric type transducers (type is switch-selectable)
Coupling:
CM type - charge amplifier with software-selectable time
constants of 0.5 and 5 seconds
VM type - ac coupling to remove dc bias voltage with high
pass response of 0.1Hz (-3dB)
INPUT CONNECTOR: Female BNC
AMPLIFIER:
Zero Temperature Stability:
±10μV/°C RTI, typical, after 30-minute warm-up
Charge Amplifier Zero Stability:
±1.2pC/°C RTI typical at 0.5 second time constant
DC Gain Accuracy & Stability:
±0.1% at +23°C; ±25ppm/°C
AC Accuracy: (typical at X2 gain step)
Input Frequency/Bandwidth: 500/3000 50/200
Spurious Free Dynamic Range: 100dB 110dB
Signal to Noise: 90dB 95dB
Signal to Distortion: 100dB 110dB
MEASUREMENT RANGES:
VM type transducers: ±10.5V, ±5.25V, ±2.1V, and ±1.05V
CM type transducers: ±200 000pC, ±100 000pC,
±40 000pC, ±20 000pC, ±10 000pC, ±4000pC, and
±2000pC
RESOLUTION: 0.0015% of range
CALIBRATION: Excitation sample
EXCITATION:
0, 1, 2, 4, 5, 10 and 20mA selections for VM type transducers
Accuracy:
±1% + (±30μA) typical at 1 to 20mA
Voltage Compliance: 0 to 28V.
Temperature Stability:
±50ppm/°C
A/D CONVERTER :
Type:
16-bit successive approximation with integrated sample
and hold
Integral Linearity Error:
±2 LSB
FILTERS:
Linear phase, analog, 6-pole anti-aliasing filter, and 256-tap,
programmable, FIR digital filter (lowpass)
Passband Frequency:
User-selectable 1Hz to 4kHz