Maintaining Disposables Maintaining the VITEK
®
2 Instrument
Figure 7-3: Cannula and Tubing Assembly
11) Place the new bag of saline on its shelf on the top of the instrument,
placing the eyelet of the bag over the post at the back of the shelf.
12) While pushing the release button (8, Figure 7-4), insert the saline
dispenser tube (9) into its housing. After the tube is fully inserted, release
the button.
Figure 7-4: Saline Dispensing Chamber
Note: Make sure that the saline dispenser tube is firmly seated in the housing and
that the tubing flows smoothly without any kinks or twisting.
13) Grasp the filter disk (10) and firmly push it onto the connector located
above the saline dispenser tube.
Avoid excessive pressure, however, as it could cause the filter to break.
14) Press the round plate (11) to allow saline to flow into the saline dispenser
tube. Do not allow the saline level to fill above the top inlet port (12,
Figure 7-5).