User manual

Maintaining the VITEK
®
2 Instrument Maintaining Disposables
VITEK
®
2 Instrument User Manual 7-7
510731-10EN1
Figure 7-5: Saline Fill Line
15) Release the round plate (11, Figure 7-4).
16) Close both the front access door and the saline access door to the
instrument.
17) Press Done.
The following message appears:
Performing Dispenser Self Test
The self test checks for proper installation of the saline dispenser tube. It
is recommended that you wait for the test to complete (about 20
seconds) before leaving the instrument.
If the test is successful, the Maintenance menu appears.
If the test fails, the following message displays:
Saline Dispenser Tube installation error
There is a problem with the saline dispenser tube installation. Inspect the
tube installation and correct it as needed.
18) Press Continue to return to the Change Saline screen.
19) At the Change Saline screen, press Cancel to return to the Maintenance
menu or Done to repeat the test.
20) Perform the Dispenser/Pipettor diagnostic test, described on page 8-21,
to ensure that the Dispenser Station is functioning properly.
12