User manual

7-8 VITEK
®
2 Instrument User Manual
510731-10EN1
Maintaining Disposables Maintaining the VITEK
®
2 Instrument
Installing an Accessory Kit, Part B: Pipette Tips
If the card transport system is processing cassettes through the Card
Transport Station, the following message appears:
Card Transport System is busy. Please try later.
CAUTION: Use aseptic technique when installing an accessory
kit.
1) Access the Change Pipette Tips function using the path:
Main Menu > Utilities > Maintenance > Change Pipette Tips
2) The following message appears:
About to change pipette tips. Do you wish to continue?
3) Press Yes to continue with the pipette tip change.
The following message appears:
Proceed with adding pipette tips.
4) Open the front access door (1, Figure 7-2) of VITEK
®
2 to expose the
Pipettor Station.
5) Open the Pipette Tip access door.
6) Rotate the pipette tip container (2, Figure 7-6) 90° in a counterclockwise
direction.
Figure 7-6: Location of Pipettor Tip Container
3
2