User manual

Maintaining the VITEK
®
2 Instrument Cleaning the Carousel
VITEK
®
2 Instrument User Manual 7-11
510731-10EN1
Figure 7-8: VITEK
®
2 XL Carousel Cleaning Screen
If cards are not processing in the selected incubator, and the carousel
sections have not been removed, the following screen displays:
Figure 7-9: VITEK
®
2 XL Preparing for Section Removal Screen
Note: When the Continue button is pressed, you are required to remove all of the
sections. This will require about four minutes. The Cancel button provided
here is the only opportunity to cancel this operation.
2) Press Continue.
The following message displays:
Remove the incubator access cover now
3) Open the Waste Collection door for the carousel being cleaned.
4) Open the top access door for the carousel being cleaned.
Note: The remaining steps may be easier if you also remove the Waste Collection
tray.