User manual
Maintaining the VITEK
®
2 Instrument Cleaning the Carousel
VITEK
®
2 Instrument User Manual 7-13
510731-10EN1
Figure 7-11: Molded Incubator Access Cover
When you remove the cover, the message on the screen changes to:
Preparing for Section Removal
6) When the carousel is in position, the following screen displays:
Figure 7-12: Carousel Cleaning Prompt
7) The carousel is removed in four sections. Grasp the first carousel section
and pull it slightly forward (toward the front of the instrument). You
should then be able to pull the section up and to the right, as shown in
Figure 7-13.
3