User manual
7-16  VITEK
®
 2 Instrument User Manual
510731-10EN1
 Cleaning the Carousel Maintaining the VITEK
®
 2 Instrument
3) Remove the incubator access cover (1, Figure 7-14) from the carousel 
by holding the retaining pin (2) on the top of the cover to the right while 
pulling the cover slightly forward and to the right, out of the instrument.
Figure 7-14: Machined Incubator Access Cover
or
Remove the incubator acces cover (3, Figure 7-15) from the carousel by 
lifting the top of the cover straight up from the instrument.
Figure 7-15: Molded Incubator Access Cover
1
2
3