Figure 7-20: Replacing the Button Memory Module
3) Place the other end of the module against the holding bracket (4) and 
push the module towards the side of the cassette.
The module should snap into place.
Cleaning the Boats
Note: The procedures in this section should be performed monthly, or as conditions 
require.
Note: Do not replace the boats if you are going to clean the inside of the instrument 
using the procedures that begin on page 7-24.
Removing Boats for Cleaning
Note: This procedure should be performed while the instrument is idle. If you are 
performing this procedure to remove only a particular boat, the system can 
continue to process with as few as one boat.
1) Access the Boat Cleaning function using the path:
Main Menu > Utilities > Maintenance > Cleaning > Boat Cleaning
The following screen appears:
BIOHAZARD WARNING
The boats should be considered as potentially contaminated 
and should be handled appropriately. Qualified laboratory 
personnel should take the usual precautions necessary for 
infectious agents.
2
3
4
1