User manual

Maintaining the VITEK
®
2 Instrument Cleaning the Boats
VITEK
®
2 Instrument User Manual 7-21
510731-10EN1
Figure 7-20: Replacing the Button Memory Module
3) Place the other end of the module against the holding bracket (4) and
push the module towards the side of the cassette.
The module should snap into place.
Cleaning the Boats
Note: The procedures in this section should be performed monthly, or as conditions
require.
Note: Do not replace the boats if you are going to clean the inside of the instrument
using the procedures that begin on page 7-24.
Removing Boats for Cleaning
Note: This procedure should be performed while the instrument is idle. If you are
performing this procedure to remove only a particular boat, the system can
continue to process with as few as one boat.
1) Access the Boat Cleaning function using the path:
Main Menu > Utilities > Maintenance > Cleaning > Boat Cleaning
The following screen appears:
BIOHAZARD WARNING
The boats should be considered as potentially contaminated
and should be handled appropriately. Qualified laboratory
personnel should take the usual precautions necessary for
infectious agents.
2
3
4
1