Cleaning the Instrument Interior Maintaining the VITEK
®
2 Instrument
The front legs (1, Figure 7-22) of the boat are positioned in the channel
of the base pan.
The boat is oriented with the arrow (2) pointing toward the back.
Figure 7-22: Positioning of Boat
6) Press the Press here after completion option button shown below after
replacing the boat.
The first screen reappears with the boat count at 1 of 4.
7) Repeat this procedure, starting at Step 2, until all four boats have been
replaced.
8) Press the Previous Screen key when the display shows:
Current number of Boats: 4 of 4.
Cleaning the Instrument Interior
This section describes the procedure for cleaning the inside of the instrument,
when the instrument has been shut down and no cards are processing.
To troubleshoot optics while cards are processing, see Cleaning Optics
(Cards Processing) on page 8-23.
The procedures in this section should be performed monthly, or as conditions
require.
BIOHAZARD WARNING
All organism suspensions should be considered as potentially
infectious. Qualified laboratory personnel should use
acceptable procedures for biohazardous material.