User manual
7-28  VITEK
®
 2 Instrument User Manual
510731-10EN1
 Cleaning the Instrument Interior Maintaining the VITEK
®
 2 Instrument
9) Replace the empty test card collection tray in the Waste Collection 
Station.
10) Close the top access door. 
11) Close the Waste Collection door.
Note: After cleaning the optics and returning power to the instrument, perform the 
Optical Diagnostic Test on page 8-26.
If you are not performing the remaining cleaning procedures in this section, 
skip to the procedure Turning the Instrument On on page 7-31.
Cleaning the Base Pan, Vacuum Seal and Vacuum Chamber
Note: Perform the instrument shutdown procedure on page 7-25 before proceeding 
with this procedure.
Note: If the boats have not already been removed, perform the procedure to 
remove them. (See page 7-21.)
The base pan, vacuum seal, and vacuum chamber are designed to withstand 
any of the following cleaning methods.
• 10% bleach solution
• Phenolic cleaning solution
Note: Remove all boats (see Removing Boats for Cleaning on page 7-21) and shut 
down the instrument (see Shutting Down the Instrument on page 7-25) before 
performing this procedure.
The base pan is the surface on which the boats move. The vacuum seal is 
the bottom surface of the vacuum chamber where the chamber makes 
contact with a boat (1, Figure 7-25).
BIOHAZARD WARNING
The base pan should not be cleaned unless the drip pan is 
properly installed. See Cleaning the Drip Pan on page 7-29 for 
instructions on cleaning the drip pan.