 Cleaning the Drip Pan Maintaining the VITEK
®
 2 Instrument
VITEK
®
 2 has a small container, called the drip pan, located underneath the 
Dispenser/Pipettor Station. The pan is designed to catch any spills that occur 
at this station.
Note: This procedure should be performed after you clean the base pan. See 
Cleaning the Base Pan, Vacuum Seal and Vacuum Chamber on page 7-28.
Removing the Drip Pan
1) Lower the bottom access door located in the bottom, front portion of the 
instrument.
The plate is held magnetically and should swing down on its hinges 
when you pull on it at both ends.
2) Grasp the handle of the drip pan (1, Figure 7-26) and carefully pull it from 
the instrument.
Figure 7-26: Drip Pan Handle
Cleaning the Drip Pan
The drip pan should be thoroughly cleaned and dried before it is replaced in 
the VITEK
®
 2 instrument. The pan material is designed to withstand any of 
the following cleaning methods:
• Automatic dishwasher with standard laboratory detergent
BIOHAZARD WARNING
The contents of the drip pan may include biohazardous 
materials. Take appropriate precautions.