User manual

7-32 VITEK
®
2 Instrument User Manual
510731-10EN1
Cleaning the SCS Maintaining the VITEK
®
2 Instrument
Cleaning the SCS
Note: This procedure should be performed monthly, or as conditions require.
The SCS base unit, keypad, display, and bar code scanner (except the clear
lens) are designed to withstand any of the following cleaning solutions when
applied first to a cleaning cloth and then to the SCS:
10% bleach solution
Phenolic cleaning solution
To clean the SCS:
1) Remove the cassette if one is on the SCS base unit.
2) Push the power switch on the SCS to the off position.
3) Remove the SCS AC power cord from its outlet.
4) Note the metal contacts (1, Figure 7-27), and take care only to wipe them
gently in an up and down motion.
CAUTION: To prevent damage to the metal contacts on the SCS
base unit, make sure that you wipe them GENTLY in an up and
down motion, rather than side to side.
Figure 7-27: SCS Base Unit Metal Contacts
BIOHAZARD WARNING
All organism suspensions should be considered as potentially
infectious. Qualified laboratory personnel should use
acceptable procedures for biohazardous material.
1