 Status Screen Error Conditions Troubleshooting the VITEK
®
 2 Instrument
6 – 05:16. Time error occurred.
Reviewing an Instrument Halted Screen
1) Press Help (?) to view the error/message queue.
Note: If there is more than one message, use the Down Arrow to review and 
address each message.
2) Note the error code number, primary error code, and secondary error 
code listed on the error/message queue.
3) If the secondary error code listed is Card Transport System Failure, 
proceed to Resolving a Transport Halted Screen to resolve the error. 
4) Look up the primary error code and then the secondary error code in 
Table 8-2 in this chapter. If an error recovery procedure is given, follow 
its directions, then continue with Step 5.
5) Press the Previous Screen key to exit the error/message queue and 
return to the Halted screen.
6) Press Go to resume processing.
• If the recovery procedure was successful, the Status screen appears 
with a Status of OK.
• If the Instrument Halted screen reappears, record the error code 
number and call bioMérieux.
Resolving a Transport Halted Screen
Note: Only cards loaded into the jammed boat will be terminated. After the jam has 
been cleared, cards loaded into other non-jammed boats will continue to 
process and will be finalized when card processing is completed.
1) Press the Previous Screen key to exit the error/message queue and 
return to the Halted screen.
2) Press Go to access the Transport Halted screen.
3) Open the front access door and the cassette load/unload door to check 
for any jams or obstructions throughout the transport system. Things to 
look for are:
• pipette tips that have fallen onto the base pan
• test cards that are not seated properly in the cassette
• boats that are not seated properly in their track