User manual

8-8 VITEK
®
2 Instrument User Manual
510731-10EN1
Status Screen Error Conditions Troubleshooting the VITEK
®
2 Instrument
indicate a Card Transport System Failure error again after the system
has rebooted.
6) Close the front access door and the cassette load/unload door.
7) Press Go to exit the Transport Halted screen.
8) The instrument will make a clicking noise and automatically reboot.
If the recovery procedure was successful, the Status screen appears
with a Status of OK once the instrument has rebooted.
If the Instrument Halted screen reappears, record the error code
number and call bioMérieux.
Message Status Screen
This status indicates that there is an error message in the error/message
queue that has not been viewed. Notice that the Status field (1, Figure 8-5)
has changed to Messages, indicating that there are error messages in the
error/message queue.
Figure 8-5: VITEK
®
2 Status Screen
Press Help (?) to view the error/message queue.
If there is more than one message or error, press the Down Arrow to review
and address each message.
LASER WARNING
All access doors and covers must remain closed when
processing cards to avoid exposure to laser light.
1