Cassette Load Processing Errors Troubleshooting the VITEK
®
2 Instrument
Figure 8-7: Bar Code Error Message
2) Open the door to the Cassette Load Station and remove the cassette.
The door should remain open while resolving the bar code error.
3) Remove the card from the slot indicated in the message.
4) Press the Previous Screen key twice to exit the error queue to the Main
Menu.
5) Access the Resolve Bar Code Errors option on the Main Menu.
The Resolve Bar Code Error screen appears (Figure 8-8).
Figure 8-8: Resolving Bar Code Errors
6) Using the numeric keypad, enter the bar code from the bar code label.
Press Enter.
BIOHAZARD WARNING
The cassette should be considered as potentially contaminated
and should be handled appropriately. Qualified laboratory
personnel should take the usual precautions necessary for
infectious agents.