User manual
Troubleshooting the VITEK
®
 2 Instrument Cassette Load Processing Errors
VITEK
®
 2 Instrument User Manual 8-15
510731-10EN1
Note: A matching identification card is one having the same accession ID and test 
type (gram +/gram -) as the susceptibility card.
During the bar code read, the instrument checks for one of these two 
conditions. If the instrument detects an inoculum error:
• The cassette is backed out to the Cassette Load Station.
• The error alarm is activated.
Resolving an Inoculum Error
1) Press the Error/Message key to display the message queue.
For an inoculum source error you see the following screen:
Figure 8-10: Inoculum Source Error Message
Note: Another inoculum error message is “Grouping Error at Slot: NN”.
2) Open the Cassette Load door and remove the cassette.
3) Place the inoculum source next to the appropriate slot.
4) Place the cassette back into the Cassette Load Station and close the 
door.
This restarts the processing cycle for the cassette.
BIOHAZARD WARNING
The cassette should be considered as potentially contaminated 
and should be handled appropriately. Qualified laboratory 
personnel should take the usual precautions necessary for 
infectious agents.