User manual

8-26 VITEK
®
2 Instrument User Manual
510731-10EN1
Boat Transport Positions Troubleshooting the VITEK
®
2 Instrument
15) Continue with the Optical Diagnostic Test, noting the conditions of the
test.
Optical Diagnostic Test
The Optical Diagnostic Test should not be performed if ID cards are
processing. However, in some cases you need to perform the optical
diagnostic test in response to an optics error.
Transmittance Optics Errors with ID Cards Processing — After
cleaning the transmittance optics in response to a transmittance optics
error, await the next read cycle of the cards to determine if the cleaning
was effective.
Transmittance Optics Errors without ID Cards Processing You may
perform the Optical Diagnostic Test after cleaning transmittance optics in
response to a transmittance optics error.
CAUTION: Performing the Optical Diagnostic Test while ID cards
are processing may result in ID cards being TERMINATED.
1) Access the Optical Diagnostic Test using the path:
Main Menu > Utilities > Diagnostics > Optics
2) (VITEK
®
2 XL Only) Select a Reader.
The following message appears for either the VITEK
®
2 or VITEK
®
2 XL:
About to perform optical test. Do you wish to
continue?
3) Press Yes to continue.
4) If the optical test fails, call bioMérieux.
5) Once the test is complete, press Previous Screen to exit the window.
Boat Transport Positions
The diagram below shows the six boat transport positions. During
troubleshooting of jams involving the card transport system, it is important to
know the boat positions inside the instrument. See Resolving a Transport