2 Instrument User Manual
510731-10EN1
Instrument Hardware Components VITEK
®
2 Instrument
Optics
The VITEK
®
2 instrument performs its identification and susceptibility
analyses by continually monitoring the growth and activity of organisms
inside the wells of the test cards. The optics system performs this function.
Transmittance Optics
The transmittance optics (1, Figure 4-10) use visible light to directly measure
organism growth. These optics are based on an initial light reading of a well
before significant growth has begun. Light transmittance samplings of the
same well every 15 minutes measure organism growth by how much light is
prevented from going through the well.
Figure 4-10: Optics
The optics use light emitting diodes (LEDs) that produce light at the
appropriate wavelengths, and silicon photodetectors to capture the
transmitted light. The system is self-calibrating.
Note: The optics should be cleaned periodically. See Cleaning Optics (Normal
Maintenance) on page 7-26.
Card Ejection
The card ejection function permanently removes test cards from the carousel
after their testing is completed. The mechanism that performs this function is
the same drive belt system that moves the test cards through the reader.
Instead of returning to the carousel, an ejected test card continues on to the
Waste Collection Station.
The amount of time that cards are held before being ejected automatically
from the carousel is set by an option in the System Configuration screen at
the workstation computer. Cards can also be ejected at any time using a
manual ejection function.