User manual
5-4  VITEK
®
 2 Instrument User Manual
510731-10EN1
 Configuration Options Configuring the VITEK
®
 2 Instrument
For information on setting this option, see the procedure Defining 
Character Sets on page 5-9.
Schedule Instrument QC Status
The QC Status is a report of the current incubator temperature and the status 
of the optics systems. The report is sent to the VITEK
®
 2 workstation where it 
is recorded with the date and time the report was taken.
The configuration option allows you to schedule up to three times of day 
when a QC report is taken and recorded at the workstation.
The Instrument Quality Control Status includes two parameters:
• Carousel Incubator Temperature
• Optics Systems
The instrument continuously monitors these parameters so their status can 
be determined at any time. The user interface provides a function, called 
Display Instrument QC Status, to access this information.
Figure 5-1: The Instrument QC Status Screen
1 — This field displays the time and date that the previous QC status was 
recorded. The field is blank if this function has never been used before.
2 — Temperature. For VITEK
®
 2, this field displays the current carousel 
temperature in degrees Celsius. For the VITEK
®
 2 XL, the temperatures for 
carousel A and B are displayed.
3 — Optical Systems. This field displays the status of the transmittance 
optical system.
4 — Record. Press this button to transmit the QC status information to the 
workstation. During the transfer the screen displays the message:
1
2
3
4