Recording QC Report
The instrument user interface contains a function with which you can 
manually record a QC Status. You can set this option if you want the report 
recorded automatically. For example, you can have the report recorded at the 
same time every day.
For information on setting this option, see the procedure Setting Time for 
QC Status on page 5-10.
Cassette Mode
The Cassette Mode tells the VITEK
®
 2 instrument whether or not you use a 
Smart Carrier Station, and is critical to instrument operation. The option 
enables or disables the instrument’s Button Memory Reader, a device that 
reads information stored in the button memory.
The Cassette Mode option defaults to Smart Carrier, meaning that an SCS is 
being used. Use this option to change the mode to Cassette Only if your 
laboratory does not have an SCS.
For information on setting this option, see the procedure Using Option 
Boxes on page 5-11.
Dilution Mode
The Dilution Mode option can be set to either Automatic or Pre-diluted. In 
the Automatic mode, the Dispenser/Pipettor station in the VITEK
®
 2 
instrument automatically prepares the inoculum for AST cards that have 
matching ID cards. In the Pre-diluted mode, the Dispenser/Pipettor station is 
disabled.
If you use an SCS, the operation of the Dispenser/Pipettor station is 
determined by the information in the cassette’s button memory.
Note: If the VITEK
®
 2 instrument is in Smart Carrier mode, the Dilution Mode option 
cannot be accessed. If you attempt to use the option, the screen displays the 
message:
This option is used with Cassette Only mode
However, if you are not using an SCS, this option must be set as follows:
• To Automatic if you want VITEK
®
 2 to prepare your AST card inocula.
• To Pre-diluted if you prepare AST card inocula manually.