User manual

Configuring the VITEK
®
2 Instrument Configuration Options
VITEK
®
2 Instrument User Manual 5-5
510731-10EN1
Recording QC Report
The instrument user interface contains a function with which you can
manually record a QC Status. You can set this option if you want the report
recorded automatically. For example, you can have the report recorded at the
same time every day.
For information on setting this option, see the procedure Setting Time for
QC Status on page 5-10.
Cassette Mode
The Cassette Mode tells the VITEK
®
2 instrument whether or not you use a
Smart Carrier Station, and is critical to instrument operation. The option
enables or disables the instrument’s Button Memory Reader, a device that
reads information stored in the button memory.
The Cassette Mode option defaults to Smart Carrier, meaning that an SCS is
being used. Use this option to change the mode to Cassette Only if your
laboratory does not have an SCS.
For information on setting this option, see the procedure Using Option
Boxes on page 5-11.
Dilution Mode
The Dilution Mode option can be set to either Automatic or Pre-diluted. In
the Automatic mode, the Dispenser/Pipettor station in the VITEK
®
2
instrument automatically prepares the inoculum for AST cards that have
matching ID cards. In the Pre-diluted mode, the Dispenser/Pipettor station is
disabled.
If you use an SCS, the operation of the Dispenser/Pipettor station is
determined by the information in the cassette’s button memory.
Note: If the VITEK
®
2 instrument is in Smart Carrier mode, the Dilution Mode option
cannot be accessed. If you attempt to use the option, the screen displays the
message:
This option is used with Cassette Only mode
However, if you are not using an SCS, this option must be set as follows:
To Automatic if you want VITEK
®
2 to prepare your AST card inocula.
To Pre-diluted if you prepare AST card inocula manually.