User manual

Processing VITEK
®
2 Test Cards Instrument Status Screen
VITEK
®
2 Instrument User Manual 6-3
510731-10EN1
Instrument Status Screen
You use the Status screen more often than any other screen in the VITEK
®
2
user interface. Since it is so basic to the workflow, the Status screen is very
easy to display and in many cases, the interface displays it automatically.
Displaying the Status Screen
There are three ways to display the Status screen:
Automatically displays at the end of the instrument’s power up initialization
process.
Automatically displays from the Main Menu whenever that menu is left
unattended for more than one minute.
Displays from the Main Menu by pressing the Previous Screen key.
Instrument Status Field
The Status field appears at the top of the VITEK
®
2 Status screen.
Figure 6-1: VITEK
®
2 Status Screen in Cassette Only Mode
The Status field (1, Figure 6-1) displays one of the following:
OK. This status means that all of the subsystems in the instrument are
working normally, and that the instrument is ready to accept test cards for
processing.
Warming. This status is seen after the instrument is turned on. It means
that the incubation temperature in the Reader Station has not reached its
specified temperature. Test cards cannot be processed until this status
changes to OK.
1
2
3
4
5
6
7