 Instrument Status Screen Processing VITEK
®
 2 Test Cards
• Messages. This status indicates that there is an error message in the 
error/message queue that has not been viewed. Press the Help (?) key to 
view the error/message queue.
• Errors. This status appears after
 the error/message queue is accessed if 
the error condition has not been resolved. This status can be cleared only 
by resolving the condition that generated the original message.
• Cleaning. This status indicates that either all four boats, or at least one 
carousel section, have been removed for cleaning. Test card processing 
cannot resume unless all carousel sections, or at least one boat, are 
replaced.
Non Standard Card Reading Mode
A large X will be displayed over the Status screen when the instrument is 
configured in a Non Standard Card Reading mode. Non Standard Card 
Reading mode is for diagnostic troubleshooting by trained bioMeriéux Field 
Service Engineers only. If the instrument is configured in a Non Standard 
Card Reading Mode, do not process any cards and contact bioMeriéux 
immediately.
Figure 6-2: VITEK
®
 2 Status Screen in Non Standard Card Reading Mode
Card Capacity
The MAX Available Slots field (2, Figure 6-1) indicates the number of 
unoccupied slots in the instrument. You can load one or more cassettes 
containing up to that number of test cards. If you load more than that number, 
some of the test cards will not be processed unless additional slots become 
available by the time the test cards reach the carousel.