Datasheet
1. L / C / R
2. Q / D / R
3. A(1KHz) / B(120Hz)
4. P(PAL) / S(SER)
5. A(AUTO) / M(MENU)
6. 0 / 1: Main Display MSD,
8: while changing range ,
9: OL
7. 6~10 : Main Display Data
8.
9.
10. LSD
11. Main Display Range
12. MSD
13. 12~15 : Second Display Data
14.
15. LSD
16. Second Display Range , 9: OL
17. Sequence 0~9 cycling
18. MSD
19. 18~21 : D value
20.
21. LSD
22. Range for D value , 9 : OL
23. MSD
24. 23~26 : Q value
25.
26. LSD
27. Range for Q value , 9 : OL
28. S(SET) / _(normal)
29. F(FUSE) / _(normal)
30. H(HOLD) / _(normal)
31. R(Present value) / M(Maximum value) /
I(Minimum value) / X(Max-Min value) /
A(Average value) / _(normal)
32. R(REL) / S(REL SET) / _(normal)
33. L(LIMITS) / _(normal)
34. T(TOL) / S(TOL SET) / _(normal)
35. B(Backlight) / _(normal)
36. A(Adapter insert) / _(normal)
37. B(Low Battery) / _(normal)
38. CR (ASCII : 0DH)
39.
nl(LF) (ASCII : 0AH)
S 0 1 2 3 4 5 6 7 8
A [ U ± 1 9 9 9 9 ]
↓
Start Code
↓
Data
Format
↓
Stop Code
↓
Setting
Value
[U?9999] : REL SET setting value
[V?9999] : Limits Hi setting value
[W?9999] : Limits Lo setting value
[X?9999] : TOL SET setting value
[Y?9999] : TOL SET Hi setting value
[Z?9999] : TOL SET Lo setting value
After receiving of setting data from PC (U/V/W/
X/Y/Z), and will send out the receiving data again
then it is convenient for PC to check the accuracy
of data output are total 7 digits.
d. Command [BXXXXXX]:
Exit the SETUP mode
RS-232 Command Table
Use IR as the interface of Data transmission, and
use external computer to start RS-232 functions.
RS-232 Interface Parameter :
Baud rate : 1200
Parity check : EVEN
Data buts : 7
Stop bits : 1
(1) Setup Selections
a. Command S: Meter will enter the Setup mode
and return "SETUP READY..x".
(x:firmware version)
b. Command
[E(L/C/R)(Q/D/R)(P/S)(A/B)(A/M)(0~6)]:
Main function change
(L/C/R): L, C or R test function
(Q/D/R): Q, D or R
(P/S): P-PAL / S-SER
(A/B): A - 1KHz / B - 120Hz
(A/M): A - AUTO / M - MANUAL
(0~6): Manual range
c. Command [A-AAAAA]: Default change
(2) Read Data
Command N: Read Meter current data and
status.
Data format: There are 39 ASCII code.
The Main Display data just send L/C/R data
that don't process before (like the data before
REL).
The Second Display data is the same as Main.
(4) RS-232 Output Chart for Second Display :
range
Q / D R R R
(Rs=100Ω)
(Rs=1K
Ω,
10K
Ω
)
(Rs=100KΩ)
1 999.9 99.99Ω 99.99Ω X
2 99.99 999.9Ω 999.9Ω 999.9Ω
3 9.999 9.999KΩ 9.999KΩ 9.999KΩ
4 .9999 99.99KΩ 99.99KΩ 99.99KΩ
5 X X 999.9KΩ 999.9KΩ
(5) FORMULA :
R
P
= R
S
(1+Q
2
)
C
P
= C
S
[1/(1+D
2
)]
C
S
= C
P
(1+D
2
)
L
P
= L
S
[1+(1/Q
2
)]
L
S
= L
P
[Q
2
/(1+Q
2
)]
range
Rs R L Rs C
1KHz/120Hz 1KHz 120Hz 1KHz 120Hz
0 100Ω 20.000Ω 2000.0µH 20.000mH 100KΩ 2000.0pF 20.000nF
1 100Ω 200.00Ω 20.000mH 200.00mH 100KΩ 20.000nF 200.00nF
2 100Ω 2000.0Ω 200.00mH 2000.0mH 10KΩ 200.00nF 2000.0nF
3 1KΩ 20.000KΩ 2000.0mH 20.000H 1KΩ 2000.0nF 20.000µF
4 10KΩ 200.00KΩ 20.000H 200.00H 100Ω 20.000µF 200.00µF
5 100KΩ 2000.0KΩ 200.00H 2000.0H 100Ω 200.00µF 2000.0µF
6 100KΩ 10.000MΩ 1000.0H 10000H 100Ω 2000.0µF 20.000mF
(3) RS-232 Output Chart for Main Display :