Digital Storage Oscilloscope 35
2.7.12 CAN Triggering
1. Press the [Trig Menu] button on the front panel to enter the Trigger system function
menu.
2. Press the Type softkey, then turn the Universal Knob to select CAN and push down
the knob to confirm.
3. Press the Source softkey, turn the Universal Knob to select CH1~CH4 as the
trigger source.
4. Set the Baud Rate.
5. Set the Idle Level.
6. Set the Identifier.
7. Press When softkey to set trigger condition.
Start - The oscilloscope triggers at the start of a frame.
Remote ID- The oscilloscope triggers on remote frames with the specified ID.
a. Press the Identifier softkey to select the ID number.
b. Press the Identifier softkey to move the current bit, and turn the universal to
select the byte to be set.
Data ID - The oscilloscope will trigger on data frames matching the specified ID
Note: The Identifier means Remote Id and Data ID.