User manual

Specifications
DSO-1000D Digital Oscilloscope User Manual 62
Chapter 8 Specifications
8.1 Technical Specifications
All specifications herein mentioned apply to the DSO-1062D oscilloscopes. Before checking an
oscilloscope from our company to see if it complies with these specifications, make sure it meets
the following conditions:
The oscilloscope must have been operating continuously for twenty minutes under the
specified operating temperature.
The Do Self Cal operation must be performed through the Utility menu if the operating
temperature changes by more than 5.
The oscilloscope must be within the factory calibration interval.
All specifications are guaranteed unless noted ‘typical’.
Oscilloscope Specifications
Horizontal
Sample Rate Range 1GSa/s at single channel, 500MSa/s at dual channel
Waveform Interpolation (sin x)/x
Record Length
Maximum 1M samples per single-channel; maximum 512K samples
per dual-channel (4K, 40K optional)
SEC/DIV Range
DSO-1062D, DSO-1102D: 4ns/div to 80s/div, in a 2,4,8 sequence
DSO-1202D: 2ns/div to 80s/div, in a 2,4,8 sequence
Sample Rate and
Delay Time Accuracy
±50ppm over any ≥1ms time interval
Delta Time
Measurement Accuracy
(Full Bandwidth)
Single-shot, Normal mode
± (1 sample interval +100ppm × reading + 0.6ns)
>16 averages
± (1 sample interval + 100ppm × reading + 0.4ns)
Sample interval = s/div ÷ 200
Position Range
DSO-1062D,DSO-1102D
20ns/div to 80μs/div (-8div × s/div) to 40ms
200μs/div to 40s/div (-8div × s/div) to 400s
DSO-1202D
2ns/div to 10ns/div (-4div × s/div) to 20ms
Vertical
A/D Converter
8-bit resolution,
each channel sampled simultaneously