4
.
Panel Operation
46
waveform is output until a trigger event is received.
Note: In Burst mode, the upper limit of the carrier frequency is half of the
maximum frequency of the original carrier. Taking a sine wave as an
example, the maximum carrier frequency is 200MHz. Press to set
the carrier to 200MHz. Then press the Burst softkey menu, then press or
Burst to display the original carrier frequency to 100MHz.
Warning:
If necessary, the burst period will increase to accommodate the specified
number of cycles.
For infinite count bursts, an external or manual trigger source start pulse train
(except internally) is required.
(6)
Press Source to select the source. The internal is the internal signal source;
the external is the external source of the Ext Trig/Burst/Fsk In interface on
the rear panel. Under the external signal source, the slope can be selected
as positive/negative. (Positive: Select the trigger signal when rising) ;
negative: select to output the trigger signal when falling); manually select
manual trigger, in the N cycle burst interface, press the Trigger under the
current channel of the front panel to output a burst.
Set the gated burst