Instructions

4.Panel Operation
26
Press the Harmonic Type softkey menu and the instrument outputs the fundamental and even
harmonics.
Odd harmonic
Press the Harmonic Type softkey menu and the instrument outputs the fundamental and odd
harmonics.
Order harmonic
Press the Harmonic Type softkey menu and the instrument outputs the fundamental and
harmonics in sequence.
customize
Press the Harmonic Type softkey to customize the number of times the harmonics are output.
The maximum number of times is 16. The 16-bit binary data is used to represent the output state
of the 16th harmonic, respectively, 1 means the output of the corresponding subharmonic is
turned on, and 0 means the output of the corresponding subharmonic is turned off. The user
only needs to use the numeric keypad to modify the value of each data bit (note: the leftmost bit
indicates the fundamental wave, fixed to X, and modification is not allowed). For example, set
16-bit data to X001 0000 0000 0001, indicating the output fundamental, 4th harmonic, and 16th
harmonic. Note: The harmonics of the actual output are limited by the currently specified
“harmonic times”.
Set the harmonic times
Press the Next soft key to enter the next page, and then press the Harmonic Times softkey to
confirm whether the “Harmonic Times” menu item is highlighted; if not, press the Harmonic
Times softkey. In parameter 5 of Figure 5-11, the parameter value of the harmonic order appears
as a blinking cursor. Use the knob or numeric keypad to set the desired value, which can be set
from 2 to 16 times.
Set the harmonic number
Press the Next soft key to enter the next page, and then press the Sequence soft key to confirm
whether the Number” menu item is highlighted; if not, press the Sequence soft key. In
parameter 6 of Figure 5-11, the parameter value of the serial number appears as a blinking cursor.
Use the knob or numeric keypad to set the desired value, which can be set from 2 to 16 times.
Set the harmonic amplitude
Press the Next soft key to enter the next page, and then press the Harmonic Amplitude softkey
to confirm whether the “Harmonic Amplitude” menu item is highlighted; if not, press the
Harmonic softkey soft key. In parameter 7 of Figure 5-11, the parameter value of the harmonic
amplitude appears as a blinking cursor. Use the knob or numeric keypad to set the desired value.
Set the harmonic phase
Press Next to enter the next page, and then press the Harmonic Phase softkey to confirm
whether the “Harmonic Phase” menu item is highlighted; if not, press the Harmonic Phase
softkey. In parameter 8 of Figure 5-11, the parameter value of the harmonic phase appears as a
blinking cursor. Use the knob or numeric keypad to set the desired value.