9. MEASURING
To receive precise measured values, the infrared thermometer must be adjusted to the ambience temperature. Let
the device adjust to the new ambience temperature after a site change.
Longer IR measurements of high temperatures at low measuring distances lead to heating of the meters and thus to
wrong measurements. For precise measured values: The higher the temperature, the larger should the measuring
distance and the shorter the measuring duration be.
Glossy surfaces will falsify the measuring result in the IR measurement. To compensate, the surface of glossy parts
can be covered with adhesive tape or matte-black paint. The devices cannot measure through transparent surfaces,
such as glass. Instead, it will measure the surface temperature of the glass.
a) Display symbols and values
In measuring operation, various parameters and symbols are displayed on the screen. The following list explains
their meaning:
IR measuring function without camera
image
IR measuring function with camera
image
Start symbol
Measurement is triggered by pushing
the measuring button
Lower alarm level activated (Low)
Upper alarm level activated (High)
Target laser is activated
Symbol for internal memory
ε Pre-set degree of emission
AT Ambient temperature
DP Dew point
DIF
Temperature difference between min
and max.
MIN
Minimum temperature of the current
measurement
Dew point measuring function
Video recording running
Pause icon
Measurement is interrupted by releas-
ing the measuring button
Lower alarm value has been undercut
Upper alarm value has been exceeded
Marks the picture centre in camera
mode
Symbol for SD card memory
TK Temperature thermocouple type K
RH% Relative humidity
WB Evaporation temperature (WetBulb)
AVG
Average temperature of the current
measurement
MIN
Maximum temperature of the current
measurement