User manual

28
You can establish the connection as illus-
trated at the front sockets or with the plug
clamps on the back (19/20).
Activate all DC outputs with the button
„OUTPUT ON“.
All output voltages and currents are con-
trolled via the „master“ device.
Sensor operation „sense“
The sensor operation „sense“ allows the precise voltage setting directly on the consumer. This reliably
compensates a possible voltage drop via the connection cables. Sensor operation is only possible in
normal (master) mode.
Switch the power unit off.
Connect the consumer with the output sockets (observe the cor-
rect polarity!) or the plug clamps (O/P) on the power unit.
Connect the sensor cable from the connection clamps of the con-
sumer with the sensor input on the power supply observing the
right polarity. The positive lead must be connected to the clamp
„RS+“ (18) and the negative lead to the clamp „RS -“ (21).
To do so, press the little slot
inwards with a screwdriver and
insert the insulated conductor into
the round opening.
Switch on the mains power supply at the power switch (1). The
voltage and current display appears after a short initiation period.
Set the parameters according to your specifications as
described in the chapter „Commissioning“.
Check once again that the output voltage was set correctly and
activate the DC output with the button „OUTPUT ON“.
Now you can switch on the connected consumer.