tiE Time setting Press the button MAX MIN button to switch
between the values in the time setup page.
The top four digits show the year setting (YYYY),
below it, you will nd the month setting and the date
setting (MMDD).
At the bottom part of the display, you will nd the time
setting. (HH:MM).
rAt Sampling rate Set the sampling rate for real-time display / recording.
You will see the current sampling rate (mm:ss) after
entering the setup page.
Press the MAX MIN button to switch between the
values in the time setup page.
Press the ▲T1·T2 button (2) or HOLD button (7)
to adjust the value.
CLR1 Clear memory Shows the amount of empty memory.
Clear the internal memory. Select YES to clear and no
to keep the memory.
TI OFFSET T1 offset value Set the offset value of measurement T1.
T2 OFFSET T2 offset value Set the offset value of measurement T2.
5. To return to the normal display mode, press and hold the SET button.
Determining any deviation of a sensors measurement accuracy
1. Turn the thermometer on and place the relevant sensor in an optional temperature calibrator or in
any other reference location with a known and stable temperature.
2. Wait until the thermometer’s display has stabilised.
3. If there is a difference, you can manually compensate it (see section “Adjusting settings”).
10. USB CONNECTION
1. Insert the mini USB plug of the mini USB cable into the USB socket (6) of the measuring instrument.
2. Insert the USB plug of the mini USB cable into a vacant USB port on a personal computer (PC).
3. Start the software “Multiple Data Logger” and follow the instructions under the “Help > Help topics”.